Ear-to-Ear Communication
ReSound
LiNX
2
supports ear-to-ear communication. The master
switch to disable/enable this functionality can be found in the main
menu under “Instrument.” The “Ear-to-Ear Communication” window
keeps a continual record of the enabled Ear-to-Ear features.
Tinnitus Sound Generator
The default setting for Tinnitus Sound Generator (TSG) is off. When
TSG is in the “Standard” mode, TSG options such as noise type,
level, modulation and modulation speed can be changed; when
TSG is in the “Nature Sounds” mode, only sound level can be
changed.
Synchronized Acceptance Manager
Synchronized Acceptance Manager allows for gradual gain
increases from current gains to final gains over a specified amount
of time. Synchronization of this feature is recommended in case of
a loss of one hearing instrument during the acclimatization time, as
this will prevent the remaining hearing instrument from progressing
more than 10% past the progression of the lost hearing instrument,
in the event it is found. Once activated, the Acceptance Manager
progress will appear on the Fit screen. Final gain levels will appear
as dotted lines. To set a final level different than current gain
prescription, ensure the box above the gain handles indicates “End
Gains (B).” Select the gain handles desired and fine tune the gains
to the desired levels.
Beeps and Volume Control
The “Beeps and Volume Control” window has two tabs: Standard
Beeps and Manual Controls. In the Standard Beeps tab, all the
acoustic indicators can be demonstrated, changed, added and
removed. Low or high frequency beeps can also be selected. The
Manual Controls tab contains options such as customizing push
button function, enabling and disabling the program button and
adjusting the volume control range.