Page 1 of 2    June 2016

Although ResMed S8 series devices operate from 12V DC the DC input ports run opposite polarity from the Freedom

TM

CPAP Battery; therefore, the use of the 150W sine wave power inverter is required in conjunction with battery. 

DO

NOT connect a ResMed S8 device to the battery without using the sine wave power inverter as doing so will
cause irreversible damage to the PAP device.

REQUIRED COMPONENTS

Freedom

TM

 CPAP Battery Kit

150W Sine Wave Power Inverter (inverter)

CONNECTION INSTRUCTIONS

Plug the DC input cord into the DC OUT jack of the battery.

Plug the inverter into the DC input cord.

Plug the AC power cord of the PAP device into the inverter.

Turn the PAP device on and it should begin functioning.

The digital display screen will show the charge level of the battery. You may press the charge level indicator
button to illuminate this screen. It will stay illuminated for a few seconds at a time.

When not in use disconnect all devices from the battery thus allowing the battery to completely shut down.

PAP AC POWER CORD

150W SINE WAVE
POWER INVERTER

DC INPUT CORD

DC OUT JACK

The inverter can also be used to charge the battery in a vehicle (vehicle does not need to be running) and to power/charge
other small electronic devices with the 

b

attery. The inverter is not proprietary to the 

Freedom

TM 

CPAP Battery and can be

used with other external batteries.

OPTIONAL FREEDOM

TM

 CPAP BATTERY CHARGING ACCESSORIES

Auto Charge DC Cable {BPS DC.CABLE}: Used to charge the battery in a vehicle (vehicle must be running).