NOT013281-8
Ventilator casing
Clean the outside of the device with a dry cloth or, if necessary, with a damp sponge.
WARNING
•
The ventilator may be disinfected using current procedures approved by your service
centre. However, users are strongly advised not to use abrasive powders or solvents.
•
If other products are used, they must comply with the conditions set by the
pharmacopoeia of your country, must be known not to leave any residue and must not
affect the operation of the ventilator.
•
Never spray any liquid directly onto the ventilator casing.
7.3 Preventative maintenance
Every 12 months, the ventilator must be serviced by a qualified technician. This service must
include a check of all the functions, calibration of the pressure and volume measurement devices
and a safety check incorporating testing of the alarms.
Preventative maintenance schedule
For details of the servicing tasks, refer to the Service Manual.
Servicing
Frequency
People qualified to perform
service
Oxygen sensor
Six months (this period will vary depending on the oxygen
concentration used and the ambient temperature)
Trained technicians
Preventative
maintenance
Yearly
Trained technicians
Preventative
maintenance
10,000 hours
Trained technicians
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...