How to use Elisée 350
NOT013281-8
The results of the test are displayed. They tell you:
•
Whether the test failed or was successful
•
What type of patient circuit is connected
•
The resistance and compliance values of the patient circuit
•
If the oxygen sensor is defective and must be replaced (if so, a flashing message appears
on the screen). Replace the sensor and repeat the manual test.
7. If the test was successful, press
to return to the main screen. If the test failed, press
to redo the test.
WARNINGS
•
If a double circuit is not connected properly while the device is performing the last step
of the test (step 6), the device may incorrectly state that a single circuit is installed. If this
happens, reconnect the circuit correctly, press
, then redo the test.
•
If the O
2
sensor is defective (the message “O2 sensor gain calculation: FAILURE” is
displayed during the manual test or if the Replace O2 Cell alarm is triggered), do not use
the device until the O
2
sensor has been replaced.
4 Testing the alarms
The operation of all of the visual and audible alarms must be checked regularly, in particular prior
to patient connection and during continuous ventilation.
See “Alarm testing procedures” on page 74.
CONTINUER
CONTINUER
Obstruct the inspirator
y
outlet !
Cancel
Cancel
Cancel
When done, press:
CONTINUE
CONTINUE
Attach
y
our circuit to the
m
achine
Leave the patient end of the circuit
open.
CONTINUE
CONTINUE
Obstruct the end of the patient
circuit.
CONTINUE
CONTINUE
When done, press:
When done, press:
4. Cover the insufflation outlet on the ventilator, then
press “Continue”. The ventilator measures the rate
of leak, if any.
5. Connect the patient circuit accessories, excluding
the mask (the other end of the patient circuit should
be left free), then press “Continue”. The ventilator
measures the inspiratory resistance at 30 L/min and
60 L/min.
6. Cover the free end of the patient circuit and press
“Continue”. The ventilator determines the type of
patient circuit and its resistance and compliance,
and calibrates the expiratory spirometry and the
oxygen sensor.
Back
Back
RESTART
RESTART
TEST
TEST
Back
Back
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...