XCell™ C410:V3 Controller
User Guide
39 repligen.com UG-C410v3-06
Table 18. Batch Set-up screen values
Field
Value
Controller Set Up
Pump model No
Select XCell ATF® 6
Control Mode
Select Flow
Slope Function Enable
Do not enable
Bioreactor pressure
0
∆
height (cm)
0
Alarm Delay (min)
1
Alarm Set Points
Value
HiHi Alarm Set Points
7 psi
LoLo Alarm Set Point
-7 psi
Alarm Delay (x10 msec)
50 for both
Pump parameters
Value
P-Pressure Offset (psi)
0.5
E-Pressure Offset (psi)
-0.5
P-Delay (%)
70
E-Delay (%)
70
P-Overtime (%)
120
E-Overtime (%)
120
P-PV Step size (%)
3
E-PV Step size (%)
3
Press accept changes.
Table 19. Advanced Set Up screen values
Field
Values
User Set point range
XCell ATF® 4
Device
XCell ATF® 6
Device
XCell ATF® 10
Legacy Device
XCell ATF® 10
Device
Max P-Flow (LPM)
10
20
80
100
Min P-Flow (LPM)
1
5
20
20
Max E-Flow (LPM)
10
20
80
100
Min E-Flow (LPM)
1
5
20
20
PV1 Max (%)
95
PV1 Min (%)
5
PV2 Max (%)
95
PV2 Min (%)
5