28 repligen.com UG-C410v3-06
Information continued from Table 9.
Button
Description
Accept Settings
This button will enable the new settings. Exiting the screen without Accepting the
settings will discard them.
Analog Input select
This button will select which of the Analog Inputs is selected.
Capture First
Calibration Point
This button will capture the raw input value for the first point for the selected
analog input.
Capture Second
Calibration Point
This button will capture the raw input value for the second point for the selected
analog input.
Actual Value
This button will finalize the two-point calibration performed for the
corresponding line.
Basic Set Up
Press to switch to Basic Set Up Screen
Advanced Set Up
Press to switch to Advanced Set Up Screen
Help Guide
Press to switch to Start Up Guide Screen
Accept/Discard change
Accept Change and Discard Change to accept or reject any parameter change on
the screen.
All Primary
All Primary Screen Buttons are displayed to navigate to those screens.
Figure 13. Trending screen
The Trending screen displays an Overview screen from where the following trends are selected.