TangenX™ SIUS™ Cassette

 

 

USER GUIDE 

Table 3

 Recommended Crossflow Rates 

 

Crossflow 

LP Screen 

4-8 L/min/m

2

 

10 psi (0.7 bar)* 

EP Screen 

6-12 L/min/m

2

 

5 psi (0.35 bar)* 

J Channel 

10-15 L/min/m

2

 

<1 psi (0.07 bar) 

* Typical ∆P measured with wat

er and permeate closed 

 

 

Table 4

 

Maximum Recommended Operating Pressures

 

Maximum Operating Pressures at 30°C 

Forward 

100 psi (7 bar) 

Reverse 

7 psi (0.48 bar) 

 

 

Table 5

 

Air Integrity Test Specification

 

Air Diffusion Rates 

Ultrafiltration 

   

1kD thru 5k

≤ 323 sccm/m

2

 @ 15 psi (1 bar) 

Ultrafiltration 

   

10kD thru 300kD

 

≤ 323 sccm/m

2

 @ 7.3 psi (0.5 bar) 

Microfiltration 

   

≥ 0.1 um

 

≤ 323 sccm/m

2

 @ 3 psi (0.2 bar) 

 

Maximum Operating Temperature

 50°C 

CHEMICAL COMPATIBILITY 

TangenX™

 SIUS

™ membrane cassettes are compatible with the following:

 

 

ACN (<15%) 

 

DMF, DMSO (<40%)

 

 

DMAC (<15%) 

 

Phosphoric acid (<1M) 

 

Sodium Hypochlorite (<400ppm)

 

 

Sodium Hydroxide (<0.5M)

 

TangenX™

 SIUS

™ membrane cassettes are NOT compatible with the 

following:

 

 

Pure aromatic and chlorinated h

ydrocarbons 

 

Ketones 

 

Polar aromatics 

 

Aliphatic esters 

A more comprehensive list is available in the cassette validation guide.

 

 

CAUTION

 

In the event that the cassette is subjected to any of the conditions listed 

below, it is recommended that you perform both cassette integrity and water 

flux tests to ensure your cassette is not damaged.  Damage may occur as a 

result of the following: 

 

Dropping on hard surfaces, or other mechanical shock.

 

 

Poking with sharp objects on screened surfaces

 

 

Excessive 

feed pressure.

 

 

Excessive permeate backpressure, or pressurizing the filtrate port.

 

 

Exposure to harsh chemicals.

 

 

Freezing.

 

 

Excessive heat. 

 

Drying out 

– 

ultrafiltration 

or microfiltration membrane that is allowed 

to dry out can permanently damage the pore structure.

 

 

MEMBRANE CASSETTES MUST REMAIN WET AT ALL TIMES TO 

MAINTAIN PRODUCT INTEGRITY AND PERFORMANCE.  

 

FOR TECHNICAL SUPPORT OR ORDER ASSISTANCE PLEASE CALL YOUR 

LOCAL SALES REPRESENTATIVE.

  

 
 

REPLIGEN TANGENX™ STANDARD WARRRANTY

 

Repligen Corporation

 warrants its TangenX™ products will meet their 

applicable published specifications when used in accordance with their 
applicable instructions for a period of one year from shipment of the products.  

REPLIGEN MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED.  THERE 
IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR 
PURPOSE.  

The warranty provided herein and the data, specifications and 

descriptions of Repligen TangenX™ products appearing in published catalogues 
and product literature may not be altered except by express written agreement 
signed by an officer of Repligen.   Representations, oral or written, which are 
inconsistent with this warranty or such publications are not authorized and if 
given, should not be relied upon. 

In the event of a breach o

f the foregoing warranty, Repligen’s sole obligation 

shall be to repair or replace, at its option, the applicable product or part 

thereof, provided the customer notifies Repligen promptly of any such breach.  

If after exercising reasonable efforts, Repligen is unable to repair or replace 

the product or part, then Repligen shall refund to the customer all monies paid 

for such applicable product or part.  

REPLIGEN SHALL NOT BE LIABLE FOR 

CONSEQUENTIAL, INCIDENTAL, SPECIAL OR ANY OTHER DAMAGES 

RESULTING FROM ECONOMIC LOSS OR PROPERTY DAMAGE SUSTAINED.  

 

Phone 508.845.6400  |  Fax 508.845.3030 

www.repligen.com/tangenx

 

[email protected] 

 

 

Отзывы: