REPLIGEN TangenX PRO Скачать руководство пользователя страница 1

IF.PUG.014     R4

 

TANGENX™ PRO CASSETTE INSTALLATION 

1.

 

Lift the end plate off the manifold of the TangenX™ PRO cassette holder. 

2.

 

Remove the protective blue sheet and rinse the Silicone gaskets with deionized 

water or WFI.  Place a rinsed gasket flat against the bottom manifold; ensure 

that the holes in the gasket line up with the holes in the manifold. 

3.

 

Using scissors carefully open the cassette bag to remove cassette. 

WARNING:  Each cassette is stored in an aqueous solution 

containing 15-20% glycerin and 0.1% sodium azide, pH 7 - 10.  Follow 

standard safety procedures for handling aqueous glycerin/sodium 

azide, including the use of gloves, safety goggles, and lab coat. 

4.

 

Place the cassette into the holder flat against the gasket.  Place another gasket 

on top of the cassette.  Ensure that the holes in the manifold, gaskets, and 

cassette are completely aligned.  If you are using multiple cassettes, continue 

the same gasket/cassette/gasket pattern, ending with a gasket between the 

last cassette and the end plate. 

5.

 

Place the end plate on top of the last gasket of the cassette or cassette stack. 

6.

 

Install the tie-rod spacers (if used) and washers on each bolt leaving a 

minimum of 18 mm (0.75 inch) of thread exposed on the rod.  By hand, screw 

the nut on each bolt and hand tighten evenly by alternating from one nut to 

the other.  Bolts must be further tightened to within the recommended torque 

values using a calibrated manual torque wrench. 

7.

 

Proceed to step 8 for a four bolt hardware design or skip to step 9 for a two 

bolt hardware design. 

8.

 

Using the calibrated torque wrench with a deep style socket, tighten each hex 

nut ¼ turn following the torque sequence illustrated in Figure 2 for a four bolt 

horizontal hardware.  In ¼ turn increments, tighten the first (H1), second (H2), 

third (H3), and fourth (H4) nut alternating back and forth until the torque 

wrench "clicks".  Repeat this sequence until the wrench "clicks" without 

turning the nut.  The "click" of the torque wrench indicates that the nut has 

reached the set point torque value. 

 

9.

 

Using the calibrated torque wrench with a deep style socket, tighten each hex 

nut ¼ turn following the torque sequence illustrated in Figure 3 for a two-bolt 

vertical hardware.  Tighten the first nut ¼ turn, and then tighten the second 

nut ¼ turn alternating back and forth until the torque wrench "clicks".  Repeat 

this sequence until the wrench "clicks" without turning the nut.  The "click" of 

the torque wrench indicates that the nut has reached the set point torque 

value. 

 

 
 
 
CAUTION:  Nuts must be 

tightened uniformly to 

avoid damaging the 

cassette.  Leakage may 

result from non-parallel 

plate alignment or over 

compression of the 

cassettes at one end.   
 
 

 
 
 
 

10.

 

Wait 5-10 minutes and allow the gaskets to relax before re-torqueing. Check 

each nut, per the Figure 2 or 3 sequences, using the torque wrench at its set 

point torque value.    

11.

 

Re-torque as needed, to create a liquid-tight seal. 

NOTE:  Torque may change during processing as the cassettes may 

compress, or as the cassettes expand or contract with temperature 

changes.  Periodically check the torque of the bolts and adjust 

torque as needed.  

 
 
 
 

 

 

 

 

 

 

 

 

 

   

 

  

 

 

 
 
 
 
 

 

 

 

TangenX™ PRO 

Cassette Holder 

COMPATIBILITY GUIDE 

REPLIGEN TANGENX™ STANDARD WARRANTY 

Repligen Corporation

 warrants its TangenX™ products will meet their 

applicable published specifications when used in accordance with their 

applicable instructions for a period of one year from shipment of the 

products.  

REPLIGEN MAKES NO OTHER WARRANTY, EXPRESSED OR 

IMPLIED.  THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR 

A PARTICULAR PURPOSE.  

The warranty provided herein and the data, 

specifications and descriptions of Repligen TangenX™ products appearing in 

published catalogues and product literature may not be altered except by 

express written agreement signed by an officer of Repligen.   

Representations, oral or written, which are inconsistent with this warranty or 

such publications are not authorized and if given, should not be relied upon. 
In the event of a breach of the foregoing warranty, Repligen’s sole obligation 

shall be to repair or replace, at its option, the applicable product or part 

thereof, provided the customer notifies Repligen promptly of any such 

breach.  If after exercising reasonable efforts, Repligen is unable to repair or 

replace the product or part, then Repligen shall refund to the customer all 

monies paid for such applicable product or part.  

REPLIGEN SHALL NOT BE 

LIABLE FOR CONSEQUENTIAL, INCIDENTAL, SPECIAL OR ANY OTHER 

DAMAGES RESULTING FROM ECONOMIC LOSS OR PROPERTY DAMAGE 

SUSTAINED.  

 

FIGURE 3 

TORQUE SEQUENCE FOR  

TANGENX™ PRO CASSETTES 

(2 Bolt Design) 

Phone 508.845.6400  |  Fax 508.845.3030 

www.repligen.com/tangenx

 

[email protected] 

 

FIGURE 2 

TORQUE SEQUENCE FOR TANGENX™ PRO CASSETTES 

(4 Bolt Design) 

Отзывы: