KrosFlo® TFDF™ Lab System
User Guide
57
repligen.com IF.UG.023 R1
Taring
Figure 47. Manual Mode Overview screen
•
Pressure Sensor Zero:
Zeros the pressure sensor reading to 0.0 psi or bar. The zero button
disappears during automation sequences to prevent accidental zeroing.
•
Scale Taring:
Tares the reading of the scales to 0.0000 kg, the tare button disappears during
automation sequences to prevent accidental zeroing, other functional items are specific to
the mode overview screen and described in their respective mode sections.
•
Flow meter Zero:
Zeros the flow meter located between the bioreactor and the filter,
zero
the flow meter only after flow path is primed only
, the zero button disappears during
automation sequences to prevent accidental zeroing
•
Reset Totalizer:
While the permeate is flowing the totalizer sums the total volume. For
accuracy, reset the totalizer before beginning a process
Select
Manual
in the Run Setpoints screen to change to manual mode. In manual mode, the TFDF™
setpoints will no longer be available. To operate in manual mode, return to the Overview Screen.
Pressure
sensor zero
Scale tare