KrosFlo® TFDF™ Lab System
User Guide
52
repligen.com IF.UG.023 R1
Figure 40. Concentration/Diafiltration/Concentration Mode Overview screen
Wizard feature
The Wizard feature enables the system to run the process automatically with preset setpoints. Users
input five parameters, and the setpoints for Concentration 1, Diafiltration step, and Concentration 2
will be automatically calculated:
•
Process volume
•
PCV (Packed Cell Volume)
•
Initial Concentration Factor (CF 1)
•
Expected Final Yield
•
Final Permeate Volume
The Wizard feature is accessed in the Run Setpoints screen when the Concentration/ Diafiltration/
Concentration mode is selected. Its calculations determine:
•
Diafiltration buffer volume needed
•
Permeate volume at diafiltration pump start
•
Permeate volume at diafiltration pump stop
•
Permeate volume at run end
•
Time for reference at diafiltration pump start
•
Time for reference at diafiltration pump stop
•
Time for reference at run end
To run the Wizard feature:
1.
Select the
Start Conc./Diaf./Conc.
button in the Run Setpoints screen.