KrosFlo® TFF System – Auxiliary Scale ACSS-60K
User Guide
23 repligen.com 420-14871-002 Rev. 00
Table 11. Save mode options
Option Description
StAb
Manual printout of stable weighing result. Upon pressing
key at the moment when the
result is unstable (no
pictogram displayed), the program first waits for the stability
condition to be met, only then the printout is carried out
nStAb
Manual printout of each weighing result. In case of unstable indication, <?> sign is displayed in
front of the 'mass frame'
rEPL
Automatic printout of the first stable weighing result above <Lo> threshold (to set <Lo> threshold
go to <Lo> parameter)
rEPLi
Automatic printout with time interval set in [min] (to set the interval go to <Int> parameter)
4.
Press
key to confirm, next return to home screen.
11.2.3
Automatic printout interval
Parameter enabling to set frequency of an automatic printout. Printout interval is set in minutes
with 1 [min] accuracy within 1 [min] – 1440 [min] range.
1.
Enter
<P3.Func>
menu and select a given working mode.
2.
Enter
<Int>
function, window for enter
time interval
value is displayed.
3.
Press
key to confirm, next return to home screen.
11.2.4
Lo threshold
<Lo>
parameter allows to configure the function of automatic operation. The next measurement is
saved only when mass indication gets below the set net value of
Lo threshold
parameter.
1.
Enter
<P3.Func>
menu and select a given working mode.
2.
Enter
<Lo>
function, window for enter
Lo threshold
value is displayed.
3.
Enter respective value and press
key to confirm, next return to home screen.
12.
Working mode – Weighing
<UUGG>
is a standard working mode to carry out the weighing operation along with record of the
result to the database.
12.1
Local settings
To go to local settings, enter
<3.1.UUGG>
submenu.
Table 12. Weighing – Local settings
Submenu
Setting
3.1.1.Acc
Working mode accessibility
3.1.2.Snn
Save mode
3.1.3.Int
Time interval
3.1.4.Lo
Lo threshold