KrosFlo® TFF System – Auxiliary Scale ACSS-60K
User Guide
17 repligen.com 420-14871-002 Rev. 00
3.
Set respective value and press
key to confirm, next go to the home screen.
6.
Communication
Communication between the scale and peripheral devices is established via RS232 port. To set the
ports go to
<P4.Conn>
submenu.
6.1
RS232 (1) port settings
1.
Enter
<P4.Conn / 4.1.rS1>
submenu.
2.
Submenu
<4.1.1.bAd>
sets the baud rate. Press
key, and rates are displayed
successively one by one:
2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s.
3.
Submenu
<4.1.1.bAd>
sets the baud rate. Press
key, and rates are displayed
successively one by one:
2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s.
4.
Set respective value and press
key to confirm, next go to the home screen.
7.
Peripheral devices
<P5.ducE>
menu contains list of devices connecting with the scale.
7.1
Computer
<5.1.PC>
submenu allows to:
•
Select port to which the computer is connected
•
Enable/disable continuous transmission
•
Set frequency of printouts for continuous transmission
7.1.1
Port
1.
Enter
<P5.1.PC / 5.1.1.Prt>
submenu.
2.
Press
key, port options are displayed successively, where:
nonE
– none,
rS1
– RS232.
3.
Set respective value and press
key to confirm, next go to the home screen.
7.1.2
Continuous transmission
1.
Enter
<P5.1.PC / 5.1.2.Cnt>
submenu.
2.
Press
key, transmission options are displayed successively, where:
nonE
– Continuous
transmission disabled,
CntA
– Continuous transmission in basic unit,
Cntb
– Continuous
transmission in current/temporary unit.
3.
Set respective value and press
key to confirm, next go to the home screen.