VIONiC RTLC20-S installation guide

12

General specifications

Scale technical specifications

Power supply

5V −5%/+10%

Typically 200 mA fully terminated

Power from a 5 Vdc supply complying with the requirements for 

SELV of standard IEC 60950-1

Ripple

200 mVpp maximum @ frequency up to 500 kHz

Temperature

Storage 

−20 °C to +70 °C

Operating 

0 °C to +70 °C

Humidity

95% relative humidity (non‑condensing) to IEC 60068‑2‑78 

Sealing

IP40

Acceleration 

(system)

Operating 

400 m/s², 3 axes

Shock 

(system)

Operating 

500 m/s², 11 ms, ½ sine, 3 axes

Vibration 

(system)

Operating

100 m/s² max @ 55 Hz to 2000 Hz, 3 axes

Mass

Readhead  8.6 g

Cable  26 g/m

Readhead cable

Single-shielded, outside diameter 4.25 ±0.25 mm

Flex life > 20 × 10

6

 cycles at 30 mm bend radius

UL recognised component 

Maximum readhead 

cable  length *

3 m

Form 

(H × W)

0.4 mm × 8 mm (including adhesive)

Pitch

 

20 µm

Accuracy 

at 20 °C 

 

(includes slope and linearity)

±5 µm/m calibration traceable to International Standards

Linearity

±2.5 µm/m achievable with 2 point error correction

Supplied lengths

20 mm to 10 m 

† 

 (> 10 m available on request)

Material

Hardened and tempered martensitic stainless steel fitted 

with a self-adhesive backing tape 

Coefficient of thermal expansion 

 

(at 20 °C)

10.1 ±0.2 µm/m/°C 

Installation temperature

15 °C to 35 °C

Datum fixing

Loctite 435 with A-9585-0028 clamp

CAUTION:

 Renishaw encoder systems have been designed to the relevant EMC standards, but must 

be correctly integrated to achieve EMC compliance. In particular, attention to shielding arrangements is 
essential.

 For lengths > 2 m 

FASTRACK

 with RTLC20 is recommended. Refer to the 

VIONiC RTLC20/FASTRACK 

linear encoder system

 installation guide (Renishaw part no. M-6195-9225).

*

 Extension cables available. Contact your local Renishaw representative for details. 

Reference mark

Limit switches

Type

Customer selected 

IN-TRAC

 reference mark, directly 

embedded into incremental track. Bi-directional position 
repeatability.

Selection

Single reference mark selection by selector magnet 
(A-9653-0143) customer positioned

L ≤100 mm

Single reference mark at scale centre

L >100 mm Reference marks at 50 mm spacing (first reference mark 

50 mm from scale end)

Repeatability

Unit of resolution repeatability (bi-directional) across full 
system rated speed and temperature ranges

Type

Magnetic actuators; with dimple triggers Q limit, without 

dimple triggers P limit  

(see ‘RTLC20-S scale installation drawing’ on page 4)

.

Trigger point

The limit output is nominally asserted when the readhead 
limit switch sensor passes the limit magnet leading edge,  

but can trigger up to 3 mm before that edge

Mounting

Customer placed at desired locations

Repeatablity

<  0.1 mm

Содержание VIONiC RTLC20-S

Страница 1: ...RSLM high accuracy linear encoder Installation guide M 6195 9108 03 B VIONiC RTLC20 S linear encoder system ...

Страница 2: ...le application 5 End covers 5 Reference mark selector and limit magnet installation 6 Datum clamp 6 VIONiC encoder system quick start guide 7 Readhead mounting and alignment 8 System calibration 9 Output signals 10 Speed 10 Electrical connections 11 Output specifications 11 General specifications 12 Scale technical specifications 12 Reference mark 12 Limit switches 12 ...

Страница 3: ...series encoder system data sheet Renishaw part no L 9517 9678 Advanced Diagnostic Tool ADTi 100 data sheet Renishaw part no L 9517 9699 Advanced Diagnostic Tool ADTi 100 and ADT View software quick start guide Renishaw part no M 6195 9321 and the Advanced Diagnostic Tool ADTi 100 and ADT View software user guide Renishaw part no M 6195 9413 These can be downloaded from our website www renishaw com...

Страница 4: ... self adhesive tape is on the outside of bend 95 relative humidity non condensing to IEC 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling N heptane CH3 CH2 5 CH3 Propan 2 ol CH3 CHOHCH3 Scale and readhead Readhead only Acetone CH3 COCH3 Methylated Spirits Chlorinated Solvents ...

Страница 5: ...on P and Q limit switch sensor position Selected IN TRAC reference mark 0 08 Roll tol 0 5 13 5 4 25 4 15 10 8 75 Offset 2 75 0 5 0 25 29 Detail A Scale reading surface 5 Mounting faces Scale thickness 0 4 including adhesive P limit magnet Reference mark selector magnet Q limit magnet Optical centreline incremental and reference mark Yaw tol 0 4 Rideheight 2 1 0 15 Pitch tol 1 2 mounting holes M2 5...

Страница 6: ...035 5 8 START See page 5 FINISH See page 5 A C End cover alignment indicator 0 5 F 0 2 100 0 05 F F axis of motion F axis of motion Nominal Q limit trigger point Ra 3 2 Optional bolted reference mark selector or limit magnets 22 18 10 3 7 Ø2 2 1 85 18 5 1 0 8 3 5 Detail A 12 1 5 4 5 3 7 Datum clamp A 9585 0028 RTLC20 S scale Detail C Detail B 15 1 IN TRAC reference mark Bolted magnet type Part num...

Страница 7: ... RTLC20 S scale installation drawing on page 4 3 Thoroughly clean and degrease the substrate using recommended solvents see Storage and handling on page 2 Allow the substrate to dry before applying scale 4 Mount the scale applicator to the readhead mounting bracket using M2 5 screws Place the shim supplied with the readhead between the applicator and substrate to set the nominal height NOTE The sc...

Страница 8: ...ed by magnetic materials in close proximity In such cases they should be held in place using an additional fillet of epoxy glue or similar along the outer edge of the magnet assembly Optional bolted reference and limit magnets are available X The reference mark selector and limit actuator locations are correct for the readhead orientation shown X External magnetic fields greater than 6mT in the vi...

Страница 9: ...ate the calibration routine The LED will single flash Blue Move the readhead along the scale at slow speed 100 mm s without passing a reference mark until the LED starts double flashing Blue The system is now calibrated and ready for use Calibration values Automatic Gain Control AGC and Automatic Offset Control AOC status are stored in readhead non volatile memory at power down NOTE If calibration...

Страница 10: ...optional Advanced Diagnostic Tool ADTi 100 A 6195 0100 and ADT View software can be used to optimise signal strength in challenging installations See www renishaw com adt for more information NOTE When re installing the readhead factory defaults should be restored see page 9 Roll 0 0 5 Yaw 0 0 4 Pitch 0 1 Rideheight 2 1 0 15 mm Green spacer Readhead set up LED status Green flashing Orange flashing...

Страница 11: ...ased X The system automatically exits the calibration routine and is ready for operation X AGC is automatically switched on once calibration is complete To switch off AGC refer to the Enabling disabling AGC section X If the LED continues double flashing after repeatedly passing the chosen reference mark it is not being detected Ensure that the correct readhead configuration is being used Readheads...

Страница 12: ...imum speed m s Minimum edge separation ns 5 µm D 1 µm X 0 5 µm Z 0 2 µm W 0 1 µm Y 50 nm H 40 nm M 25 nm P 20 nm I 10 nm O 5 nm Q 2 5 nm R 50 12 12 12 7 25 3 63 1 81 1 45 0 906 0 725 0 363 0 181 0 091 25 3 40 12 12 12 5 80 2 90 1 45 1 16 0 725 0 580 0 290 0 145 0 073 31 8 25 12 12 9 06 3 63 1 81 0 906 0 725 0 453 0 363 0 181 0 091 0 045 51 2 20 12 12 8 06 3 22 1 61 0 806 0 645 0 403 0 322 0 161 0 ...

Страница 13: ... isolator Remote CAL operation CAL 0 V Remote operation of the CAL AGC is possible via CAL signal Output specifications Digital output signals Form Square wave differential line driver to EIA RS422A except limits P and Q Incremental 2 channels A and B in quadrature 90 phase shifted Signal period P Resolution S A B Z Reference Synchronised pulse Z duration as resolution Bi directionally repeatable ...

Страница 14: ...C Installation temperature 15 C to 35 C Datum fixing Loctite 435 with A 9585 0028 clamp CAUTION Renishaw encoder systems have been designed to the relevant EMC standards but must be correctly integrated to achieve EMC compliance In particular attention to shielding arrangements is essential For lengths 2 m FASTRACK with RTLC20 is recommended Refer to the VIONiC RTLC20 FASTRACK linear encoder syste...

Страница 15: ...hts reserved Renishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countries apply innovation and names and designations of other Renishaw products and technologies are trade marks of Renishaw plc or its subsidiaries Loctite is a registered trade mark of ...

Отзывы: