Slide RTLC20 scale into the 

FASTRACK

 carrier 

ensuring the scale is fed under the projections 
as shown.
Scale can be installed manually by either 
pulling or pushing it through the 

FASTRACK

 carrier. 

Alternatively use the optional 
scale installation tool 
(A-9589-0420) as shown, for 
easy installation. 

VIONiC RTLC20/

FASTRACK

 installation guide

6

RTLC20/

FASTRACK

 carrier installation

 

(Continued)

6

Ensure a gap > 1 mm

7

Optional: fix self‑adhesive end covers 

ensuring a gap of at least 1 mm between 
the end of the scale and the end cover.

Projections

NOTE:

 For instructions on how to use the scale 

installation tool, download the 

RTL scale installation 

tool (A-9589-0420) 

user guide (Renishaw 

part no. M-9589-9101) from the website at 

www.renishaw.com/encoderinstallationguides

IMPORTANT:

 If manually installing 

the scale using fingers, suitable 

gloves should be worn to protect 
against injury from sharp edges.

8

Clean 

FASTRACK

 and scale 

using a lint-free cloth.

Scale datum

The datum clamp fixes the RTLC20 scale rigidly to the substrate at the location chosen.

CAUTION:

 The metrology of the system may be compromised if the datum clamp is not used. 

The datum clamp does not need to be fitted adjacent to a reference mark. It can be positioned anywhere 
along the axis depending upon the customer’s requirements. 

Mechanical clamp 

Adhesive clamp

Using dispensing tip P-TL50-0209 apply Loctite 435 
between the scale and the 

FASTRACK

 carrier so it 

wicks underneath adjacent to user selected datum 
location as shown.

NOTE:

 Drawing shows the scale datum adjacent to 

the chosen reference mark.

NOTE:

 When the reference mark and 

datum clamp are aligned as shown, 

the reference output will be positionally 

repeatable with respect to the substrate.

Reference mark 

selector magnet

3

Do not place

 clamp tip 

over shaded area

IN-TRAC

 reference mark

12

25

Bolted datum clamp 

(A-9589-0077)

M3 × 6 low head screw, 

supplied with clamp

15 ±1

† 

Additional screws available (pack of 25 A-9584-2047).

NOTE:

 Only apply Loctite 435 in 

these gaps to ensure best bond. 
Loctite 435 will wick under the 
scale to lock it to the substrate. 

16

12

2.3

1.7

5 min 

thread 

depth

Содержание Vionic RTLC20/FASTRACK

Страница 1: ...VIONiC RTLC20 FASTRACK linear encoder system Installation guide M 6195 9225 03 A...

Страница 2: ...installation 5 Reference mark selector and limit magnet installation 7 VIONiC encoder system quick start guide 8 Readhead mounting and alignment 9 System calibration 10 Output signals 11 Speed 11 Ele...

Страница 3: ...5386081 US7550710 CN101300463 EP1946048 JP5017275 US7624513 CN101310165 EP1957943 US7839296 CN102057256 EP2294363 JP5475759 JP5755299 KR1550483 US8141265 CN1314511 EP1469969 JP5002559 US8987633 US8466...

Страница 4: ...sive tape is on the outside of the bend radius Storage and handling 70 C 20 C 95 relative humidity non condensing to BS EN 60068 2 78 2013 Humidity Storage 70 C 0 C Operating Scale and readhead Readhe...

Страница 5: ...Static bend radius 35 23 11 5 6 min Set up LED A 0 6 29 Detail A 4 2 5 0 2 5 Reference mark selector magnet FASTRACK carrier thickness 0 4 including adhesive Forward direction of readhead relative to...

Страница 6: ...A 9653 0143 Dimensions as P limit Ra 3 2 Q limit magnet A 9653 0139 Dimensions as P limit 1 5 Optional end cover Pair A 9589 0058 18 5 B A Do not cut FASTRACK carrier over shaded areas 18 0 1 15 1 P...

Страница 7: ...RACK carrier or scale Ensure the block is in the correct orientation as shown Whilst holding the block in place in a smooth motion pull down the lever to cut through the FASTRACK carrier or scale Befo...

Страница 8: ...um The datum clamp fixes the RTLC20 scale rigidly to the substrate at the location chosen CAUTION The metrology of the system may be compromised if the datum clamp is not used The datum clamp does not...

Страница 9: ...elow and aligned with the outer edge of the FASTRACK carrier Limit magnets can be positioned at any user defined location along the FASTRACK carrier but the reference mark selector magnet should be po...

Страница 10: ...o initiate the calibration routine The LED will single flash Blue Move the readhead along the scale at slow speed 100 mm s without passing a reference mark until the LED starts double flashing Blue Th...

Страница 11: ...head mounting and alignment Mounting brackets The bracket must have a flat mounting surface and should provide adjustment to enable conformance to the installation tolerances allow adjustment to the r...

Страница 12: ...d X The system automatically exits the calibration routine and is ready for operation X AGC is automatically switched on once calibration is complete To switch off AGC refer to the Enabling disabling...

Страница 13: ...z Maximum speed m s Minimum edge separation ns 5 m D 1 m X 0 5 m Z 0 2 m W 0 1 m Y 50 nm H 40 nm M 25 nm P 20 nm I 10 nm O 5 nm Q 2 5 nm R 50 12 12 12 7 25 3 63 1 81 1 45 0 906 0 725 0 363 0 181 0 091...

Страница 14: ...opto isolator Remote CAL operation CAL 0 V Remote operation of the CAL AGC is possible via CAL signal Output specifications Digital output signals Form Square wave differential line driver to EIA RS42...

Страница 15: ...al specifications FASTRACK technical specifications Form H W 0 2 mm 8 mm Pitch 20 m Accuracy at 20 C includes slope and linearity 5 m m calibration traceable to International Standards Linearity 2 5 m...

Страница 16: ...enishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other co...

Отзывы: