TONiC RKLC20-S linear installation guide

14

System calibration

Calibration is an essential operation that completes readhead set-up, with the optimum incremental and reference mark signal settings stored in the readhead’s non-volatile memory. 

Before system calibration

:

u

 

Clean the scale and readhead optical window (contamination around the reference mark may result in reference mark dephasing).

u

 

If reinstalling, restore factory defaults 

('Restoring factory defaults', page 15)

.

u

 

Ensure Automatic Gain Control (AGC) is switched off (CAL LED on readhead is not illuminated)

u

 

Maximise the signal strength along full axis of travel.

 NOTE: 

CAL routine maximum speed: < 100 mm/s or less than the readhead maximum speed, whichever is slowest (all Ti /  TD interface models). TD interface can be calibrated in either resolution.

Step 1 – Incremental signal calibration

u

 

Press the CAL button on the end of the interface for <  2 seconds using a 2 mm allen key or similar tool.

 

WARNING!

 Activating the CAL switch only requires 2.5 N force. Applying excess force may permanently damage the switch.

u

  The CAL LED will now periodically single‑flash to indicate that it is in incremental signal calibration routine.

u

  Move the readhead along the axis, ensuring you do not pass the selected reference mark until the CAL LED starts double‑flashing.

 

This indicates the incremental signal is now calibrated and the new settings are stored in the readhead memory.

u

  The system is now ready for reference mark phasing.

u

  For systems without reference mark, go to ‘Calibration routine - manual exit ’

u

  If the system does not automatically enter the reference mark phasing stage (no double‑flashing of the CAL LED) the calibration of 

 

the incremental signals has failed. After ensuring failure is not due to overspeed (> 100 mm/s, or exceeding the readhead maximum speed), 

 

exit the calibration routine, restore factory defaults 

('Restoring factory defaults', page 15)

 and check the readhead installation and system  

cleanliness before repeating the calibration routine.

Step 2 – Reference mark phasing

u

  Move the readhead back and forth over the selected reference mark until the CAL LED stops flashing and remains off. The reference mark is now phased.

 

NOTE: 

Only the chosen reference mark that has been used in the calibration routine is guaranteed to remain phased.

u

  The system automatically exits the CAL routine and is ready for operation.

u

  If the CAL LED continues double‑flashing after repeatedly passing the chosen reference mark 

it is not being detected.

 

  ‑ Ensure that the correct readhead configuration is being used. Readheads can either output all reference marks or only output a reference mark where a reference selector magnet is fitted 

 

    depending on the options chosen when ordering.

 

  ‑ Check reference mark selector magnet is fitted in the correct location relative to readhead orientation 

('RKLC20-S scale installation drawing', page 5)

.

Calibration routine manual exit

u

  To exit the calibration routine at any stage, press the CAL button. The CAL button will stop flashing.

CAL LED

Settings stored

Single flashing

None, restore factory defaults and recalibrate

Double flashing

Incremental only

Off (auto-complete)

Incremental and reference mark

Содержание TONiC T103x RKLC20-S

Страница 1: ...RSLM high accuracy linear encoder TONiC T103x RKLC20 S linear encoder system Installation guide M 9653 9568 02 A...

Страница 2: ...d clamps 9 Reference mark selector and limit magnet installation 10 TONiC quick start guide 11 System connection 12 Readhead mounting and alignment 13 System calibration 14 Restoring factory defaults...

Страница 3: ...iance Compliance This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept...

Страница 4: ...elf adhesive tape is on the outside of bend 95 relative humidity non condensing to IEC 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling Readhead only Acetone CH3 COCH3 Methylated Sp...

Страница 5: ...75 Recommended mounting faces Scale thickness 0 15 including adhesive P limit magnet Reference mark selector magnet Q limit magnet Yaw tol 0 4 Rideheight 2 1 0 15 Pitch tol 1 2 off mounting holes M2...

Страница 6: ...omatic Gain Control AGC enable disable Push and hold during power Off On cycle Restore factory defaults Refer to readhead LED functionality chart for CAL LED indications Dimensions and tolerances in m...

Страница 7: ...L 40 ML L 55 with dual limits 0 5 F 0 2 100 20 A P limit magnet A 9653 0138 Dimensions as Q limit RKLC20 S scale Reference mark selector magnet A 9653 0143 Dimensions as Q limit 35 Q limit magnet A 96...

Страница 8: ...dhesive A 9531 0342 u Appropriate cleaning solvents Storage and handling page 2 u 2 M2 5 screws u Green spacer supplied with TONiC readhead Optional parts u Renishaw scale wipes A 9523 4040 u Lint fre...

Страница 9: ...ough the guillotine as shown and place guillotine press block down onto the scale Ensure the block is in the correct orientation as shown Whilst holding the block in place in a smooth motion pull down...

Страница 10: ...nted either way round to enable easiest orientation for scale installation 5 Move axis to scale start position leaving enough room for the scale to be inserted through the applicator as shown below 6...

Страница 11: ...f the end clamp 3 The end clamp features two small regions of contact adhesive These will temporarily hold the end clamp in position while the adhesive cures Remove the backing tape from either side 4...

Страница 12: ...d reference sensors Limit trigger point The limit output is nominally asserted when the readhead limit switch sensor passes the limit magnet leading edge but can trigger up to 3 mm before that edge RK...

Страница 13: ...Green interface ideally Blue Purple CALIBRATION Press and release the CAL button on the interface The CAL LED on the readhead will be single flashing Move the readhead along the scale at slow speed 1...

Страница 14: ...5 hex head screws u Taking care not to touch the pins plug the connector into the socket in the interface ensuring correct orientation as shown u Refit the cover plate ensuring the cable ferrule is lo...

Страница 15: ...flash Normal phasing Orange flash Acceptable phasing Red flash Poor phasing clean scale and recalibrate if required CAL Operating On Automatic Gain Control On Off Automatic Gain Control Off Calibratio...

Страница 16: ...ithout reference mark go to Calibration routine manual exit u If the system does not automatically enter the reference mark phasing stage no double flashing of the CAL LED the calibration of the incre...

Страница 17: ...ting that the factory defaults have been restored u Release CAL button u Check the Readhead mounting and alignment page 13 and recalibrate the system System calibration page 14 NOTE System must be re...

Страница 18: ...Black Set up VX Clear Calibrate CAL Orange Shield Inner shield Green Yellow Outer shield Outer screen Readhead output Output connector for all interfaces 15 way D type plug Becomes alarm E for Ti opt...

Страница 19: ...0 2 70 1 35 0 540 0 270 0 135 0 054 0 027 25 10 10 8 10 3 24 1 62 0 810 0 324 0 162 0 081 0 032 0 016 20 10 10 6 75 2 70 1 35 0 675 0 270 0 135 0 068 0 027 0 013 12 10 9 4 50 1 80 0 900 0 450 0 180 0...

Страница 20: ...noise issues NOTE Maximum cable length between readhead and Ti TD interface is 10 m Single ended alarm signal termination Ti options A B C D Customer electronics Interface 5 V 4k7 4k7 1k8 100R 100nF...

Страница 21: ...Active low P Q Length of limit actuator Incremental 2 channels A and B in quadrature 90 phase shifted Reference NOTE Select standard or wide reference at time of ordering to match the requirements of...

Страница 22: ...level VX is nominal 3 3 V 0 7 to 1 35 Vpp with Green LED indication readhead and 120R termination Differential signals V0 and V0 centred on 1 65 V NOTE Ti0000A00V centred on 2 5 V Output specificatio...

Страница 23: ...o IEC 60068 2 78 Sealing readhead IP40 interface IP20 Acceleration readhead Operating 500 m s2 3 axes Shock system Operating 500 m s2 11 ms sine 3 axes Vibration readhead Operating 100 m s2 55 Hz to 2...

Страница 24: ...mark directly embedded into incremental track Bi directional position repeatability Selection Single reference mark selection by selector magnet A 9653 0143 customer positioned L 100 mm Single refere...

Страница 25: ...ershire GL12 8JR United Kingdom T 44 0 1453 524524 F 44 0 1453 524901 E uk renishaw com www renishaw com For worldwide contact details visit www renishaw com contact Renishaw plc Registered in England...

Отзывы: