TONiC RKLC20-S linear installation guide

11

System calibration

Calibration is an essential operation that completes readhead set-up, with the optimum incremental  
and reference mark signal settings stored in the readhead’s non-volatile memory. 

Before system calibration:

u

 

Clean the scale and readhead optical window (contamination around the reference mark  

  may result in reference mark dephasing).

u

 

If re-installing restore factory defaults.

u

 

Ensure Automatic Gain Control (AGC) is switched off (CAL LED on readhead is not illuminated) 

u

 

Maximise the signal strength along full axis of travel.

NOTE: CAL routine maximum speed <100 mm/s (all Ti /  TD interface models). 
 

TD interface can be calibrated in either resolution.

Step 1 – Incremental signal calibration

u

 

Press the CAL button on the end of the interface for < 2 seconds  
using a 2 mm allen key or similar tool.  
WARNING! Activating the CAL switch only requires 2.5 N force.  
Applying excess force may permanently damage the switch.

u

 The CAL LED will now periodically single‑flash to indicate that 

 

it is in incremental signal calibration routine.

u

 Move the readhead along the axis, ensuring you do not pass  

the selected reference mark until the CAL LED starts  

double‑flashing. This indicates the incremental signal is now 

 

calibrated and the new settings are stored in the  
readhead memory. 

u

 The system is now ready for reference mark phasing.

u

 For systems without reference mark, go to ‘Calibration routine - manual exit ’

u

 If the system does not automatically enter the reference mark phasing stage (no double‑flashing of the 

CAL LED) the calibration of the incremental signals has failed. After ensuring failure is not due to overspeed 
(>100 mm/s), exit the calibration routine, restore factory defaults and check the readhead installation and 
system cleanliness before repeating the calibration routine.

Step 2 – Reference mark phasing

u

 Move the readhead back and forth over the selected reference mark until the CAL LED stops flashing 

 

and remains off. The reference mark is now phased.  
NOTE: Only the chosen reference mark that has been used in the calibration routine is guaranteed to 
remain phased.

u

 The system automatically exits the CAL routine and is ready for operation.

u

 If the CAL LED continues double‑flashing after repeatedly passing the chosen reference mark 

it is not 

being detected.

  ‑ Ensure that the correct readhead configuration is being used. Readheads can either output all reference

 

      marks or only output a reference mark where a reference selector magnet is fitted depending on the

 

      options chosen when ordering.

  ‑ Check reference mark selector magnet is fitted in the correct location relative to readhead orientation 

 

      

(see ‘RKLC20-S scale installation drawing’).

CAL LED

Settings stored

Single flashing

None, restore factory defaults and recalibrate

Double flashing

Incremental only

Off (auto-complete)

Incremental and reference mark

Restoring factory defaults

When re-installing the system or in the case of continued calibration failure,  
factory defaults should be restored. 

To restore factory defaults:

u

  Switch system off.

u

  Press and hold the CAL button whilst switching the system on. The CAL LED on the readhead will flash 

several times, indicating that the factory defaults have been restored. 

u

  Release CAL button.

u

  Check the ‘Readhead mounting / installation’ and recalibrate the system.

NOTE: System must be re-calibrated after restoring factory defaults. 

Enabling /disabling  AGC

AGC can be switched on or off via the interface.

u

  Press and hold the CAL button on the interface for >3 seconds to switch AGC on or off. The CAL LED on 

the readhead will be illuminated when AGC is active.

NOTE: The system must be calibrated before switching AGC on.

Calibration routine manual exit

u

 To exit the calibration routine at any stage, press the CAL button. The CAL button will stop flashing.

Содержание TONiC RKLC20-S

Страница 1: ...RSLM high accuracy linear encoder TONiC T103x RKLC20 S linear encoder system Installation guide M 9653 9568 01 A...

Страница 2: ...allation drawing 5 Scale application 6 End clamps 6 Reference mark selector and limit magnet installation 7 TONiC quick start guide 8 System connection 9 Readhead mounting and alignment 10 System cali...

Страница 3: ...n which case the user will be required to correct the interference at his own expense NOTE This unit was tested with shielded cables on the peripheral devices Shielded cables must be used with the uni...

Страница 4: ...ive tape is on the outside of bend 95 relative humidity non condensing to EN 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling Readhead only Acetone CH3 COCH3 Methylated Spirits Chlo...

Страница 5: ...Recommended mounting faces Scale thickness 0 15 including adhesive P limit magnet Reference mark selector magnet Q limit magnet Yaw tol 0 4 Rideheight 2 1 0 15 Pitch tol 1 2 off mounting holes M2 5 t...

Страница 6: ...isable Push and hold during power Off On cycle Restore factory defaults Refer to readhead LED functionality chart for CAL LED indications 15 1 8 9 16 4 40 UNC 2 40 6 min 62 67 CAL AGC push button swit...

Страница 7: ...A 9653 0291 P limit A 9653 0292 START see page 6 Scale length L Measuring length ML L 40 ML L 55 with dual limits 0 5 F 0 2 100 20 A P limit magnet A 9653 0138 Dimensions as Q limit RKLC20 S scale Ref...

Страница 8: ...travel ensuring the backing paper is pulled manually from the scale and does not catch under the applicator Remove applicator carefully Apply firm finger pressure via a clean lint free cloth along th...

Страница 9: ...the scale NOTE Reference and limit magnets may creep when influenced by magnetic materials in close proximity In such cases they should be held in place using an additional fillet of epoxy glue or sim...

Страница 10: ...dhead Green interface ideally Blue Purple CALIBRATION Press and release the CAL button on the interface The CAL LED on the readhead will be single flashing Move the readhead along the scale at slow sp...

Страница 11: ...5 hex head screws 2 Taking care not to touch the pins plug the connector into the socket in the interface ensuring correct orientation as shown 3 Refit the cover plate ensuring the cable ferrule is lo...

Страница 12: ...iguration Also some configurations do not output overspeed alarm See product nomenclature for details Momentary status only while fault condition remains Alarm may result in axis position error re dat...

Страница 13: ...g the calibration routine Step 2 Reference mark phasing u Move the readhead back and forth over the selected reference mark until the CAL LED stops flashing and remains off The reference mark is now p...

Страница 14: ...shield on UHV cables Interface output digital Ti0004 to Ti20KD and TD4000 to TD0040 Interface Ti0004 Ti20KD TD4000 TD0040 Function Output type Signal Pin Pin Power 5 V 7 8 7 8 0 V 2 9 2 9 Incremental...

Страница 15: ...100R 100nF E Extension cable maximum 50 m dependent upon clocked output option Remote CAL operation Analogue versions only CAL 0 V All Ti TD interfaces include a push button switch to enable CAL AGC...

Страница 16: ...e high or Active low P Q Repeatability 0 1 mm Length of limit actuator Incremental 2 channels A and B in quadrature 90 phase shifted Reference NOTE Select standard or wide reference at time of orderin...

Страница 17: ...ly integrated to achieve EMC compliance In particular attention to shielding arrangements is essential RKLC20 S scale specifications Form H W 0 15 mm 6 mm including adhesive Pitch 20 m Accuracy at 20...

Страница 18: ...e marks of Renishaw plc in the United Kingdom and other countries apply innovation and names and designations of other Renishaw products and technologies are trade marks of Renishaw plc or its subsidi...

Отзывы: