TONiC RESM20 angle encoder system installation guide
General specifications
Power supply
5 V ±10%
Readhead only < 100 mA
T20x1 with Ti0000 < 100 mA
T20x1 with Ti0004 – Ti20KD or TD4000 – TD0040 < 200 mA
NOTE:
Current consumption figures refer to unterminated systems.
For digital outputs a further 25 mA per channel pair (eg A+, A−) will be drawn when terminated with 120R.
For analogue outputs a further 20 mA in total will be drawn when terminated with 120R.
Power from a 5 Vdc supply complying with the requirements for SELV of standard IEC 60950-1.
Ripple 200 mVpp
maximum
@ frequency up to 500 kHz
Temperature
(System)
Storage
−20 °C to +70 °C
Operating 0 °C to +70 °C
(UHV readhead)
B120 °C
Humidity
(System)
95% relative humidity (non-condensing) to EN 60068-2-78
Sealing
(Readhead)
IP40
(Interface)
IP20
Acceleration
(Readhead)
Operating 500 m/s
2
, 3 axes
Shock
(System)
Operating 500 m/s
2
, 11 ms, ½ sine, 3 axes
Vibration
(System)
Operating 100 m/s
2
, 55 Hz to 2000 Hz, 3 axes
Mass
(Readhead)
10 g
(Interface)
100 g
(Cable)
26 g/m
(UHV cable)
14 g/m
Readhead cable
(Standard)
Double shielded, outside diameter 4.25 ±0.25 mm
Flex life > 2
0 ×10
6
cycles at 20 mm bend radius
UL recognised component
(UHV)
Tin coated braided single screen FEP core insulation
Maximum cable length
Readhead to interface 10 m
Interface to controller
CAUTION:
Renishaw encoder systems have been designed to the relevant EMC standards, but must be correctly integrated to achieve EMC compliance.
In particular, attention to shielding arrangements is essential.
Clocked output option (MHz)
Maximum cable length (m)
40 to 50
25
< 40
50
analogue
50
RESM20 ring specifications
Material
303/304 stainless steel
Coefficient of thermal expansion
(at 20 °C)
15.5 ±0.5 µm/m/°C