RMP60 installation guide
3.10
System installation
Calibrating in a ring gauge or on a datum sphere
Calibrating a probe either in a ring gauge or on a datum sphere with a known diameter automatically
stores one or more value for the radius of the stylus ball. The stored values are then used
automatically by the measuring cycles to give the true size of the feature. The values are also used to
give true positions of single surface features.
NOTE:
The stored radius values are based on the true electronic trigger points. These values are
different from the physical sizes.
Calibrating the probe length
Calibrating a probe on a known reference surface determines the length of the probe, based on the
electronic trigger point. The stored value for length is different from the physical length of the probe
assembly. Additionally, the operation can automatically compensate for machine and fixture height
errors by adjusting the probe length value that is stored.