![Renfert SILENT TS Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/renfert/silent-ts/silent-ts_frequently-asked-questions-manual_1441140005.webp)
EN
- 1 -
FAQ - English
For information on operating elements, see
For chapter indications, refer to the complete instruction manual on the internet.
Error
Possible cause
Corrective action
Dust extractor stops
suddenly, for no appar-
ent reason.
• Dust extractor continued to be used,
despite “Change Filter” indicator.
• Unit fuse blown.
• Motor protection switch (
) has
been activated due to overload (tem-
perature, electricity).
• Motor over-temperature protector
switch activated.
• Plug on grey motor wire out of its sock-
et or loose.
• Suction turbine fault.
• In-house fuse blown.
• Check and replace the dust bag and fine particle
filter (Chapter 3.2 and 3.3).
• Check and reinstall the fuse.
• Check the motor protection switch, if necessary
reset.
• Change the motor if repeatedly activated.
• Check dust bag and filter, replace as required.
(Sec. 3.2 and 3.3).
• Check exhaust air filter, replace as required.
(Sec. 3.3.2).
• Check whether suction hose is blocked. Remove
obstruction as required.
• Dust extractor can be restarted after being allowed
to cool off for approx. 1 hour.
• Check plug (
).
• Carry out self-diagnosis (chapt. 3.5), at fault „E2“
exchange suction turbine (see spare parts list).
• Replace the in-house fuse.
• Check the total power consumption of the dust
extractor plus dust-generating devices.
Dust extractor stops
suddenly, for no appar-
ent reason.
E4 is indi-
cated on the display.
• The electronics may be overheated
due to excessively high ambient tem-
perature, e.g., if the unit is installed in
a cabinet.
• Allow the extractor to cool down.
• Attempt to improve the ventilation conditions.
Refer also to Sec. 1.1.
After switching on the
unit, the display imme-
diately indicates E4.
• Temperature sensor fault.
• After overheating, the device has not
yet cooled down sufficiently.
• Have the device repaired.
• Disconnect the device and let it continue to cool
down.
The self-test returns a
suction turbine error,
even though the tur-
bine is running.
• Filter full.
• Suction hose blocked.
• Suction turbine fault.
• Check dust bag and fine particle filter, replace as
required. (Sec. 3.2 and 3.3.1).
• Remove obstruction from the suction hose.
• Replace the suction turbine (Sec.3.6).
Insufficient extraction
force.
• Extraction force level too low.
• Filter blocked.
• Suction hose blocked.
• Dust drawer seal not tight.
• Upper front panel fails to close
properly.
• Suction hose damaged.
• Increase the extraction level (Sec. 2.3).
• Check dust bag and fine particle filter, replace as
required.(Sec. 3.2 and 3.3.1).
• Remove obstruction from the suction hose.
• Check the proper seat of the dust drawer
(Sec. 3.2).
• Check the dust drawer seal for damage and
replace as required (Sec. 3.1).
• Check the proper seat of the upper front panel
(Sec. 3.3.1).
• Check the upper front panel seal for damage and
replace as required (Sec. 3.1).
• Inspect the suction hose and replace as required.
Содержание SILENT TS