4.4 Hygienic Preparation
Please note:
⇒
Before using the SYMPRO system for the first time, prepare the components hygienically according to
the following instructions�
⇒
Fill the cleaning beaker (2, Fig� 1) with 75 g cleaning needles (equivalent to one pack)�
⇒
Fill the SYMPRO mini-cup (16, Fig�1) with ~ 5 g cleaning needles�
Before handling the disinfectant, the cleaning object or the cleaning device system components,
take care to ensure of personal safety in accordance with national practice, laboratory and hy-
giene regulations.
Manual disinfection:
D1
Disinfection process – lid & seal
Open the cleaning beaker� Remove the silicone seal from
the lid� Thoroughly rinse both objects (cover and seal) under
running water� Wipe the entire lid (outer and inner surfaces
including the thread) and the silicone seal (outer and inner
surface) with a wipe disinfectant (*)�
(*) Recommendation:
CaviWipes, Metrex
D2
Fill the cleaning beaker with disinfection liquid
Remove the cleaning beaker magnets (not applicable for first
time use)�
Fill the cleaning bowl (incl� cleaning needles) with an instru-
ment disinfection fluid for heavy organic contamination (*).
For fill quantity, see Application Table.
(*) Recommendation:
Sporicidin, Contec / Metricide OPA plus, Metrex
D3
Disinfection process – Cleaning the needles & inside the
beaker
Replace the disinfected silicone seal back into the lid or in
the case of the SYMPRO mini-cup, back into the cup�
Over the control panel set a running time of at least 10 min�
at 2000 rpm (*) and start the disinfection process�
(*) These values apply to the disinfection fluids recommend
-
ed in D2 and in the Application Table�
Other settings may be required for other types of liquid or
concentrations�
D4
Disposal of disinfection liquid
After the end of the program, remove and open the clean-
ing beaker, keep the cleaning needles with the help of the
magnet, on the bottom of the cleaning beaker, pour out the
disinfectant and dispose of according to the manufacturer’s
instructions�
D5
Disinfection process – Cleaning the outside of the bea
-
ker and & sound protection ring
Remove magnets� Remove the silicone sound-proof ring /
adaptor ring from the cleaning bowl� Wipe the surface of the
cleaning bowl (incl� thread and sound-proof ring / adapter
ring) with a standard wipe disinfectant (*)�
(*) Recommendation:
CaviWipes, Metrex
- 7 -
Содержание 65000000
Страница 1: ...KO JA ZH UK DA PL RU TR PT ES IT FR EN DE Made in Germany SYMPRO 21 7190 27022019...
Страница 2: ...KO JA ZH UK DA PL RU TR PT ES IT FR EN DE...
Страница 3: ...DE ORIGINALBEDIENUNGSANLEITUNG Made in Germany SYMPRO 21 7190 27022019...
Страница 19: ...EN TRANSLATION OF THE ORIGINAL INSTRUCTIONS FOR USE Made in Germany SYMPRO 21 7190 27022019...
Страница 35: ...FR TRADUCTION DU MODE D EMPLOI D ORIGINE Made in Germany SYMPRO 21 7190 27022019...
Страница 51: ...IT TRADUZIONE DEL MANUALE DI ISTRUZIONI ORIGINALE Made in Germany SYMPRO 21 7190 27022019...
Страница 67: ...ES TRADUCCI N DEL MANUAL DE INSTRUCCIONES ORIGINAL Made in Germany SYMPRO 21 7190 27022019...
Страница 83: ...PT TRADU O DO MANUAL DE UTILIZA O ORIGINAL Made in Germany SYMPRO 21 7190 27022019...
Страница 99: ...TR ORIJINAL KULLANIM KILAVUZUNUN EVIRISI Made in Germany SYMPRO 21 7190 27022019...
Страница 115: ...RU Made in Germany SYMPRO 21 7190 27022019...
Страница 117: ...RU 1 SYMPRO SYMPRO 1 1 2002 96 EG WEEE 2 2 1 2 2 Renfert GmbH 2 3 2000 3...
Страница 124: ...RU 5 2 4 5 5 3 9 P 1 0 30 8 P 1 7 P 1 4 5 10 5 4 10 00 10 30 P 2 10 11 10...
Страница 129: ...RU 9 Renfert SYMPRO 3 RENFERT 10 10 1 10 2 15...
Страница 131: ...PL T UMACZENIE ORYGINALNEJ INSTRUKCJI OBS UGI Made in Germany SYMPRO 21 7190 27022019...
Страница 147: ...DA OVERS TTELSE AF DEN ORIGINALE BETJENINGSVEJLEDNING Made in Germany SYMPRO 21 7190 27022019...
Страница 163: ...UK Made in Germany SYMPRO 21 7190 27022019...
Страница 165: ...UK 1 SYMPRO 1 1 2002 96 EG WEEE 2 2 1 2 2 Renfert GmbH 3...
Страница 172: ...UK 5 2 4 5 5 3 9 1 0 30 8 1 7 1 4 5 10 5 4 10 11 00 11 30 2 10 11 10...
Страница 175: ...UK 6 6 5 7 6 1 4 4 SYMPRO 5 6 7 SYMPRO Nicoclean 121 C 121 C 134 C 6 8 14 1 6 9 www renfert com p918 65000000 7 8 13...
Страница 177: ...UK 9 Renfert 3 RENFERT 10 10 1 10 1 1 15...
Страница 179: ...ZH Made in Germany SYMPRO 21 7190 27022019...
Страница 182: ...ZH 2 4 20 60 C 4 140 F 80 2 5 2 5 1 DGUV 3 VDE 0701 0702 REACH SVHC www renfert com 2 5 2 PVC 50cm 2 6 2 7 4...
Страница 187: ...ZH 5 2 4 5 5 3 9 1 0 30 8 1 7 1 4 5 10 5 4 10 11 00 11 30 5 5 2 10 11 3 9...
Страница 190: ...ZH 6 8 14 1 6 9 www renfert com p918 65000000 7 12 1 6 8 5 4 6 5 6 6 4 5 6 1 2000 100ml 5 6 6 12...
Страница 193: ...JA Made in Germany SYMPRO 21 7190 27022019...
Страница 196: ...JA 2 4 20 60 C 80 2 5 2 5 1 DGUV 3 VDE 0701 0702 2 5 2 Renfert 50 cm REACH SVHC www renfert com 2 6 2 7 Renfert GmbH 4...
Страница 201: ...JA 5 2 4 5 5 3 9 1 0 30 8 1 7 1 4 5 10 5 4 10 11 00 11 30 5 5 2 10 11 3 9...
Страница 202: ...JA 5 6 5 10 5 7 2000ml SYMPROfluid 2000 ml PP 6 6 1 MAX SYMPRO 3 6 2 4 4 D1 D7 D8 4 10...
Страница 204: ...JA 6 7 SYMPRO 121 C 121 134 6 8 14 1 6 9 www renfert com p918 65000000 7 12 1 6 8 5 4 6 5 6 6 4 5 12...
Страница 207: ...KO Made in Germany SYMPRO 21 7190 27022019...
Страница 215: ...KO 5 2 4 5 5 3 RPM 9 1 0 30 8 1 7 1 7 4 5 RPM 10 5 4 10 RPM 00 11 30 5 5 2 10 11 3 9...