  Tono-Pen

®

 XL

 

Applanation Tonometer

Quick Reference Guide

Transducer

Button

LCD Display

Keep the Tono-Pen covered with an Ocu-Film +  

tip cover at all times, even in storage. 

980041 Rev. D, 02-08-2017

Calibration   must   be   successfully   

performed:

• 

Every day before the first patient 

measurement

•  When batteries have been replaced

•  After a 

[bAd]

 calibration

•  Whenever the LCD  displays 

[CAL]

Calibration can be performed with or 

without an Ocu-Film + tip cover.

1.  Hold the Tono-Pen with the transducer end pointed straight down toward the floor.

2.  Press the button TWICE within 1.5 seconds. The instrument will “beep” and 

the LCD will display 

[CAL]

.

3.  Hold the Tono-Pen steady for approximately 15 seconds until the device 

“beeps” and the LCD displays

 [UP]

4.  Quickly and smoothly flip the Tono-Pen up, so the transducer is pointing straight up.

5.  A properly functioning Tono-Pen will display

 [Good] 

and the device will 

“beep”. Wait for the “beep” before touching the operating button again. 

6.  If 

[bAd]

 is displayed, repeat this calibration procedure.

Calibration Procedure

Ocu-Film + Instructions

CORRECT

Proper set up of the Ocu-Film

® 

+ tip cover is essential for the precision and 

accuracy of the tonometry measurement. A new Ocu-Film + tip cover must be 

used for each new patient. Be certain that the film is flat across the tip, but not 

too taut or too loose.

Position the Ocu-Film +  

applicator on the transducer.  

Roll the Ocu-Film + tip 

cover onto the tip. 

Remove the 

applicator. 

Taking an IOP Measurement

To take a measurement, press the button once.  

[8888]

 will briefly be displayed, 

followed by a double row of dashes

  [= = = =]

, indicating the Tono-Pen is ready 

to measure.  
The corneal surface needs only to be momentarily contacted.

 Indentation or 

excessive pressure is not required and may lead to injury to the eye.

 

CORRECT

INCORRECT