background image

DEF-Chek

Diesel Exhaust Fluid Tester

User Guide

Sample Well

Liquid Crystal 

Display

Scale Key

Read Key

Calibration Key

Battery Door

INTRODUCTION:

We at Reichert Analytical Instruments would like to extend our ap-

preciation to you for purchasing this instrument. It is our mission to 

provide a quality product at a very competitive price and we feel this 

instrument will prove its value time and again to you.

FEATURES:

The DEF-Chek® measures the urea content in diesel exhaust fluid 

(DEF), also referred to as AdBlue™ in Europe. The DEF-Chek®

provides results that are precise to ± 0.1% by weight.

CLEANING:

Cleaning of the measurement surface and well should be performed 

immediately after each sample reading. Never immerse the 

instrument in any liquid. When the measurement surface and well 

have been completely cleaned no residue should be present.

To properly clean the sample well and glass measurement surface 

use a mild soap and water solution or Isopropyl Alcohol followed by 

a distilled water rinse and then thoroughly dry with a soft lint and 

residue free cloth or a product such as Kimwipes®.

To clean the refractometer’s body use a soft cloth with a mild 

cleaning solution like window cleaner applied to the cloth and follow 

by wiping the instrument dry. 

CAUTION:

1. 

 Due to the nature of any automotive fluids, to avoid possible 

injury, Reichert recommends that safety glasses and gloves be worn 

when testing any sample type. 

SIMPLE SET UP:

1.

 Insert the included batteries with positive side facing outward.

2.

 Align small mark found on the battery cover with mark found on 

    the instrument body and insert cover as shown above.

3.

 Rotate the cover in the clock wise direction 

    until it is in the position shown to the right.

SAMPLE ANALYSIS:

Before analyzing a sample, it should be noted that to attain the 

greatest accuracy the refractometer should be calibrated at the same 

temperature as the envirnoment that it will be used in. As an 

example if the environment that the instrument is being used in 

changes by five degrees, then the instrument should be re-calibrated.

1.

 Inspect the glass measuring surface and well to be sure that there   

    is no residue remaining from a previous sample analysis. If the 

    measuring surface and well does not appear to be clean, then     

    refer to the cleaning section of this user guide before proceeding. 

2.

 Extract the sample to be tested and apply adequate sample to  

    the sample well to completely cover the glass measuring surface.  

    Minimum sample size is 0.3 mL.

3.

 Allow the sample time to reach the same temperature as the  

    instrument. Failure to do so will result in inaccurate measurement.

4.

 Press the “READ” key and note the displayed value. 

CLEANING PRECAUTIONS:

1. 

Never use any harsh cleaning agents that will damage the 

    instrument.

2.

 Strong solvents should never be used. Such chemicals will 

    deteriorate the prism seal and attack individual component parts  

    of the instrument which will adversely affect or destroy the 

    instrument.

AUTOMATIC TEMPERATURE COMPENSATION:

All Reichert refractometers are automatically temperature 

compensated assuring extremely accurate results. Automatic 

temperature compensation corrects readings over a range of 

temperatures. As an example; samples taken within the working 

temperature range of the instrument are corrected to a reference 

temperature of 68° F (20° C). Temperature compensation is essential 

because refractive index varies inversely with temperature. This 

feature is used in lieu of temperature control of the measuring 

surface and sample. For this feature to be effective however, it is 

necessary to allow the sample to temperature equilibrate to the 

ambient temperature of the measuring surface. Without allowing for 

temperature equilibration of the sample the read values will not be as 

accurate as they could be since the sample will be in a state of 

temperature and refractive index change.

CALIBRATION:

1. 

Be sure that the measuring surface and well are clean as outlined  

    under the “Cleaning” section.

2.

 Apply distilled water to the measurement surface.

3.

 Allow time for the distilled water sample to temperature equilibrate  

    to that of the instrument.

4.

 Press and hold the “CAL” key until CAL is displayed. While still  

    holding the “CAL” key press the “READ” key and then release both  

    keys.

5.

 A series of dashed lines will appear and successful calibration will   

    be indicated when the word END is displayed.

DISPLAYING ACTUAL TEMPERATURE:

To display the temperature at the measuring surface, momentarily 

press the “CAL” button while the sample value is displayed.  

®

Отзывы: