Menu Window Details
Key
Sub-Key
Window Number
Function
SET
Quick
Start
Step 1:
Mounting
Site
S1
Window which provides instructions on selecting a mounting site:
- Pipe is full
- Straight pipe: 10D upstream, 5D downstream
- Vertical pipe: flow must go up.
- Horizontal pipe: transducer must be on the side of pipe.
Step 2: Set
Pipe
Parameters
S2a
Window for entering the outer diameter of the pipe
Valid range: 0 to 6000mm.
S2b
Window for entering pipe wall thickness
S2c
Window for selecting pipe material
Standard pipe materials (no need to enter the material sound speed) include:
(1) carbon steel (2) stainless steel (3) cast iron (4) ductile iron
(5) copper (6) PVC (7) aluminum (8) asbestos
(9) fiberglass
S2d
Window for selecting the liner material. Select none(1) for pipes without any
liner. Standard liner materials (no need to enter liner sound speed) include:
(2) Tar Epoxy (3) Rubber (4) Mortar (5) Polypropylene
(6) Polystryol (7)Polystyrene (8) Polyester (9) Polyethylene
(10) Ebonite (11) Teflon
S3: Set Fluid
Parameters
S3a
Window for selecting fluid type
For standard liquids (no need to enter liquid sound speed) include:
(1) Water (2) Sea Water (3) Kerosene (4) Gasoline
(5) Fuel oil (6) Crude Oil (7) Propane at -45C
(8) Butane at 0C (9)Other liquids (10) Diesel Oil (11)Caster Oil
(12)Peanut Oil (13) #90 Gasoline (14) #93 Gasoline (15) Alcohol
(16) Hot water at 125C
S4: Set
Transducer
Parameters
S4a
Window for selecting transducer type
There are 30 different types of transducers for selection.
The user needs to configure the 3 transducer parameters.
S4b
Window for selecting the transducer mounting methods
Four methods can be selected:
(0) V-method (1) Z-method (2) N-method (3) W-method
S4c
Window to set the flow rate scale factor. The factory default is ‘1’.
Keep this value as ‘1’ when no calibration has been made.
S5: Install
Transducer
Now
S5a
Window for transducer mount distance and installation:
a) Mark 2 spots of the same side on pipe surface, with the distance equal to
the number shown. b) Sand & clean the spots, make surface smooth. c) put
compound on transducer surface. d) Clamp transducer onto pipe. Cable
towards outside. e) Finish wiring then power up.
S6: Verify
Window to verify the signal strength, signal quality and transit time ratio shown
S7: Diagnose
Window to Diagnose Triplet
Set Pipe
S21: Set OD
Window for entering the outer diameter of the pipe. Valid range: 0 to 6000mm.
S22: Set
Wall TH
Window for entering the pipe wall thickness.
S23: Set ID
Window for entering the inner diameter of the pipe. If pipe outer diameter and wall
thickness are entered correctly, the inner diameter will be calculated automatically,
thus no need to change anything in this window.
S24: Pipe
Material
Window for selecting pipe material
Standard pipe materials (no need to enter the material sound speed) include:
Содержание RH40
Страница 62: ...62...
Страница 63: ...63...
Страница 64: ...64...
Страница 65: ...65...
Страница 66: ...66...
Страница 67: ...67...
Страница 68: ...68...
Страница 69: ...69...
Страница 70: ...70...
Страница 71: ...71...
Страница 72: ...72...
Страница 73: ...73...
Страница 74: ...74 8 4 Sound Speed Tables...
Страница 75: ...75...
Страница 76: ...76...
Страница 77: ...77...
Страница 78: ...78...
Страница 79: ...79...
Страница 80: ...80...
Страница 81: ...81...
Страница 82: ...82...
Страница 83: ...83...
Страница 84: ...84...
Страница 85: ...85...
Страница 86: ...86...
Страница 87: ...87...
Страница 88: ...88...
Страница 89: ...89...
Страница 90: ...90...