Operating Instructions
Power ON/OFF
To turn the meter on, press and hold the
button. To turn the meter
off, press and hold the
button for approx. 3 seconds. The device will
power off automatically after idling for 180 seconds.
Backlight
The backlight will turn on after pressing any button, and turn off after 15
seconds of idling to save battery life.
Calibration Procedure
The R8005 includes a self-calibration mode to ensure measurement
accuracy.
1. While the meter is powered off, press and hold the
button
while turning the meter on.
2. Release both buttons when "CAL" and a flashing number
appear on the display indicating that the meter is now in
self-calibration mode.
3. Use the button and the button to adjust error value. The
adjustment range is -9 to 9mm.
For example, if the meter appears to be measuring 2mm lower
than the actual distance, press the to increase the calibration
value by 2mm. If the meter appears to be measuring 3mm above
the actual distance, press the button to decrease the calibration
value by 3mm.
4. Press the
button to save the set calibration result and resume
normal operation.
continued...