Страница 1: ...Loop Calibrator Instruction Manual R5820 REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Страница 2: ... 8 Output Operation Procedure 9 Current Output 9 25 Step Current Output 9 Current Output Set for Zero Point Full Scale 10 Auto Ramp 10 Simulating Transmitter Output XMT 11 Measuring Function 11 Measuring DC Current 11 12 Measuring DC Voltage 12 Providing 24V Power Supply for Measuring Loop Current 12 Battery Replacement 13 Fuse Replacement 13 Applications 13 Accessories and Replacement Parts 14 co...

Страница 3: ...itional fee for this service will apply Safety Never attempt to repair or modify your instrument Dismantling your product other than for the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center To prevent the user and the instrument from electric shock and other hazards it is nec...

Страница 4: ...Leads Alligator Clips Protective Holster Batteries Specifications Output Functions Applicable range from 64 4 to 82 4 F 18 to 28 C within one year of calibration Output Range Output Range Resolution Accuracy Note Current 20mA 0 000 to 22 000mA 0 001mA 0 05 of set value 4μA Max 1KΩ at 20mA see Note 1 Simu transmitter absorp current 20mA 0 000 to 22 000mA 0 001mA Max 1KΩ at 20mA see Note 2 Loop powe...

Страница 5: ...load is 700Ω at 20mA Note 2 Power supply range 5 to 25VDC Note 3 Temperature coefficient 0 005 of the range per C for the temperature ranges 41 to 64 4 F 5 C to 18 C and 82 4 to 104 F 28 C to 40 C General Specifications Display 5 Digit LCD Kick Stand Yes Battery Life Approx 20 hours Power Supply 2 AA Batteries Auto shut off Yes after 15 minutes off Low Battery Indicator Yes Replaceable Test Leads ...

Страница 6: ...tch Button 4 25 Single Step Setting Auto Ramp Button 5 0 100 of Setting Auto Ramp Start Button 6 Output Input Button 7 mA V Select Button 8 Output Digit Selection Button 9 Output Value Setting Button 10 Input Output Terminals REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Страница 7: ...e denoting that the zero point or the full scale point is now in calibration Indicates battery power is low and batteries need to be replaced Indicates which output digit is being set Indicates the output ramp type V mA Indicates the current output value unit of measure Indicates that the output signal is on or off REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Страница 8: ...the display 5 Press the button to toggle between AP ON AP OF AP OF indicates that the automatic power off function is disabled while AP ON indicates the automatic power off function is enabled 6 Press the 100 START button to save the required setting 7 Press and hold the POWER button to exit the maintenance state and turn the instrument off Output Function The output terminal of the instrument can...

Страница 9: ...to select the output digits 5 Press the button to change the numerical value of the set digits Hold the button and the value will keep varying 25 Step Current Output 1 Connect the test leads to the meter as stated in the Current Output procedures 2 Press the button and OUTPUT will appear on the display indicating that the calibrator is in an output state 3 Press the 25 RAMP button and will appear ...

Страница 10: ...TPUT OFF S and 4mA will appear on the display indicating that the calibrator is entering RAMP mode 4 Press the 25 RAMP button to change the type of the output ramp from S F and S indicates a low speed ramp which is set to a cycle of 60 seconds F indicates a high speed ramp which is set to a cycle of 30 seconds indicates auto stair step ramp which pauses 5 seconds at each step 5 Press the 100 START...

Страница 11: ...nal injury During operation do not apply a voltage or current exceeding the measuring range to the input terminal This could cause damage to the calibrator During the operation turn off the instrument before connecting the calibrator Connecting to the Calibrator without it being shut it down may cause damage Measuring DC Current 1 Insert one end of the test lead into the mA input terminal and conn...

Страница 12: ...t the calibrator is in an input state 3 Press the mA V button 3 times to select the V function indicated by V on the display 4 The unit refreshes twice per second If the measured value exceeds the measuring range the display will indicate the symbol OL overload Providing 24V Power Supply for Measuring Loop Current 1 Insert the test lead into the LOOP mA input jacks of the calibrator as shown in th...

Страница 13: ...unit s contact points and snap into place 5 Tighten the screw to secure the battery compartment door Fuse Replacement 1 Remove the test leads from the meter and turn the meter OFF 2 Take off the protective boot remove the four screws by using a standard blade screwdriver and then take off the cover 3 Replace the blown fuse s 4 Reinstall the cover 5 Reinstall the meter s protective boot Application...

Страница 14: ...arantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and maintenance REED Instruments total liability is limited to repair or repl...

Страница 15: ...EED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly prohibited without prior wri...

Страница 16: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...