3.3  Charging the battery 

For an optimum lifetime, you should charge the battery as often as possible. If the battery is not charged 

regularly, it will be damaged due to self-discharge. Frequent deep discharge reduces the battery life. 
During charging, the RPP10 must be connected to a control box. 
To charge the battery, do the following: 

1.  Connect the RPP10 to a control box, see section 3.2. 
2.  Connect the battery charger to the RPP10 charger jack and to a wall socket.  
3.  Check that the LED on the charger is red, indicating charging. 
4.  The charging time is approximately 9 hours.  

When the charging is completed, the LED on the charger is green. 

 

Only use the battery charger delivered together with the RPP10. 

Charging must take place in a room with good air ventilation.  

The ambient temperature must not be below 0°C or above 40°C. 

 

If the RPP10 has not been used for a long time, it is recommended to charge the 

battery for 24 hours. 

3.4  Checking battery level 

To check the battery charge level, you must connect the RPP10 to a RCB11 or RCB12 control box. For 

more information, refer to the user manual for the control box. 

Содержание RPP10

Страница 1: ...www reac se USER MANUAL RPP10 REAC POWER PACK...

Страница 2: ...cation of advanced motion systems REAC s power solutions contain compact and strong electrical actuators lift and tilt systems control units and hand controls We know that our customers have different...

Страница 3: ...SAFETY FUNCTIONS 3 3 OPERATION 4 3 1 BEFORE FIRST USE 4 3 2 CONNECTING THE CONTROL BOX 4 3 3 CHARGING THE BATTERY 5 3 4 CHECKING BATTERY LEVEL 5 4 MAINTENANCE 6 4 1 SERVICE INTERVALS 6 4 2 CLEANING AN...

Страница 4: ...and RHC15 1 2 Symbols The following symbols are used in the manual Injury to persons can occur if these instructions are not followed Damage to the product can occur if these instructions are not foll...

Страница 5: ...s used for connection of the battery charger VPI connector The VPI connector is used for connection of the control box 2 3 Battery disconnection function To protect the battery from deep discharge and...

Страница 6: ...attery and also allow moisture to come out Do not block or cover the ventilation channels at any time 2 4 2 Disabled motor activation while charging If the battery charger is connected the control box...

Страница 7: ...d to a wall socket 3 Check that the LED on the charger is red indicating charging 4 Keep the charger connected for 24 hours 3 2 Connecting the control box You connect the control box to the VPI connec...

Страница 8: ...all socket 3 Check that the LED on the charger is red indicating charging 4 The charging time is approximately 9 hours When the charging is completed the LED on the charger is green Only use the batte...

Страница 9: ...re the cable that connects the RPP10 to the control box is firmly affixed Battery Before use make sure at least 1 green battery indicator LED is lit on the hand control 4 2 Cleaning and disinfection C...

Страница 10: ...nd control drop quickly after full charge Service is needed No Charging Charger cable not connected The LED on the charger is not indicating charging Make sure the cable is properly connected to the R...

Страница 11: ...Atmospheric pressure 700 to 1060 hPa General Oxygen rich environment No Label reading environment 50 cm 500 lx Do not use the product outside the specified limits Personal injury and damage to the pro...

Страница 12: ...lectronic Equipment WEEE Directive 2002 96 CE the symbol indicates that the product must be taken to a proper disposal site and cannot be discarded in normal household waste RPP10 consist of several p...

Страница 13: ...XX Ingress of particles first character and water second character as per EN60529 IEC 60417 5840 Patient part of type B IEC 60417 5957 For indoor use only WEEE compliant CE label attached based on Low...

Страница 14: ...REAC control system products Manufacturer manufacturing date Do not use if package is damaged Fragile handle with care Keep dry Temperature limits Relative humidity limits Atmospheric pressure limits...

Страница 15: ...always is up to date nor can we take responsibility for any misinterpretation of our written context Technical specification might change due to load and external circumstances REAC products shall be...

Отзывы: