Non-Contact Thermometer
V1.1
REMOTE CARE PARTNERS
PAGE 11
Notes on Taking A Measurement
To avoid inaccurate readings, please follow these
instructions.
•
Only use this device in temperature ranges of
50°F-104°F.
•
Do not use in bright, sunny environments.
•
Take 3 measurements to confirm the results.
•
Always make sure the thermometer is in BODY
mode before taking a measurement.
The light next
to the word BODY should be lit to indicate it is in
BODY MODE. If the light next to the word SURFACE
is lit, press the "MODE" button to select BODY
MODE.
•
Store the thermometer in the room where you take
your measurements
. If you move the thermometer
to a new room to take a measurement and the
temperature in the new room is very different from
where the thermometer was stored, the reading
may not be accurate. Bring the patient to the
thermometer, don’t bring the thermometer to the
patient.
MEMORY FUNCTION
1.
Pressing "Power/Measure" and the last
reading can be viewed in the display
window.
2.
Pressing
repeatedly the “MEM” to
toggle between saved readings.
3.
Press “MODE once to switch from the
Body and Surface readings.