background image

RBR#0008819revA

 –  

15

       3. Set the battery end cap off to the side.
       4. Now you will have unobstructed access to the oil reservoir (red cap). To remove the reservoir turn the cap 

counterclockwise at least one rotation and then pull.

                        

       5. The oil absorbent towel will be contaminated with the buffer (Dow Corning FS 1265 Fluid 300 cst) and should be 

disposed of in accordance with local regulations.

                        

        6. Replace the O-ring on the waste reservoir.
        7. Replace the absorbent towel by rolling one side of the towel and inserting it into the centre of the reservoir and 

wrap the remaining towel around the reservoir in a clockwise manner (when viewed as below). The towel should only 

wrap around the reservoir once, any additional layers will make it difficult to insert. 

                       

        8. While holding the towel so that it does not unravel, reinsert the reservoir back into the instrument, and close by 
turning the cap clockwise approximately one full rotation or until tight. 

6.4.3  Repairs

We support all instruments and software that we manufacture. First line support is always available by phone or e-mail. 
Please contact us immediately if you are experiencing problems with your RBR product. It is very important to contact 

RBRquartz BPR|zero - service diagram

The complete RBR

quartz

3

 BPR|zero service drawing is available in 

Appendix B - RBRquartzBPR|zero Service 

Drawing

Содержание quartz3 BPR zero

Страница 1: ...rbr global com INSTRUMENT GUIDE BPR zero...

Страница 2: ...9 6 2 Changing the batteries RBRquartz BPR zero 11 6 2 1 Steps 11 6 3 Installing desiccant RBRquartz BPR zero 12 6 4 Maintenance and repairs 12 6 4 1 O rings RBRquartz BPR zero 12 Inspecting O rings 1...

Страница 3: ...0008819revA 3 2 Introduction You can access the RBR instrument guide on the USB data stick provided when you purchase a logger from the Help menu in Ruskin and on the RBR web site at www rbr global c...

Страница 4: ...RBR 0008819revA 4 3 Revision history Revision No Release Date Notes A 04 May 2020 Original...

Страница 5: ...by us The liability of RBR Ltd extends only to the replacement cost of the instrument The customer will bear all costs of shipment to us for repair all other costs including return shipment will be b...

Страница 6: ...rements are critical The RBRquartz BPR zero may be supplied with external power from RBRfermata power or RBRcervata power and memory canisters or cabled to an observatory for external power and realti...

Страница 7: ...35 C The instrument has 8 X AA batteries installed however this is intended only for purposes of sustaining the instrument during external power interruptions of a few minutes While the Paros will con...

Страница 8: ...2000 4000 7000 dbar Initial accuracy 0 01 FS full scale Resolution 10ppb at 16Hz sampling rate Accelerometer optional Range 3g Resolution 100ng The full mechanical drawing is available in Appendix A...

Страница 9: ...shown below Only the battery end cap needs to be opened to access the data connector batteries desiccant and waste reservoir To open the logger battery end cap the two M6 screws will need to be remov...

Страница 10: ...il the battery end cap is mated fully to the housing Remember to keep the O rings clean and avoid scratching the O ring mating surfaces Carefully inspect the O rings as described in the Inspecting the...

Страница 11: ...ent during external power interruptions of a few minutes While the Paros will continue to run and be measured the valve will not operate 6 2 1 Steps Remove the battery end cap from the logger see Open...

Страница 12: ...desiccant capsules can be opened the desiccant removed and replaced or dried 6 4 Maintenance and repairs 6 4 1 O rings RBRquartz BPR zero The single most important item of maintenance on any RBR subme...

Страница 13: ...bishment please contact us for advice Replacing an O ring Lever the O ring from its groove Use a soft plastic or wood tool do not use a metal screwdriver or any other tool which may scratch the surfac...

Страница 14: ...easure the pressure inside the housing of the instrument which is monitored by a barometer in order to correct for the drift of the sensors over time This actuation of this valve results in a small vo...

Страница 15: ...towel around the reservoir in a clockwise manner when viewed as below The towel should only wrap around the reservoir once any additional layers will make it difficult to insert 8 While holding the to...

Страница 16: ...closely as possible to the current market value based on the original purchase price less depreciation Please contact RBR if you have any questions about the instrument value Send the unit using your...

Страница 17: ...RBR 0008819revA 17 7 Appendix 7 1 Appendix A RBRquartzBPR zero Product Drawing...

Страница 18: ...RBR 0008819revA 18...

Страница 19: ...RBR 0008819revA 19 7 2 Appendix B RBRquartzBPR zero Service Drawing...

Страница 20: ...RBR 0008819revA 20 7 3 Appendix C RBRquartzBPR zero Connector Details...

Отзывы: