3.2 DO sensor (OxyGuard) care and maintenance
Storage
Store the OxyGuard dissolved oxygen sensor in the dedicated storage cap to minimize fluid loss. Storage caps are
provided with the instrument. Contact
if a replacement is needed.
O-ring
The red O-ring of the OxyGuard sensor serves two
purposes:
To retain the electrolyte during storage
To balance pressure during deployments
There are two positions for O-ring on the OxyGuard
sensor, "Transport" and "Measurement".
During transportation or storage, move the red O-ring of
the Oxyguard sensor to the "Transport" position, closing
off the port on the side of the cell.
Before deployment, move the O-ring to the
"Measurement" position to maintain the pressure
balance.
After deployment, return the O-ring to the "Transport"
position.
Support kit
RBR offers an OxyGuard sensor support kit that includes:
Membrane tool
Electrolyte solution (250ml)
Fast response membranes
Replacement O-rings
Oxyguard Support Kit and Refurbishment Guide
Check the state of your DO sensor before deployment. Look for any damage to the membrane, cloudiness of the
electrode, and buildup on the anode. If you find any damage, refurbish and re-calibrate the sensor.
Refer to Oxyguard Support Kit and Refurbishment Guide, included with the support kit, for instructions on refurbishing
your sensor. See
for instructions on calibration.