RaySafe i3 Installation & service manual – Settings explained
CONFIGURATION TAB
Figure 19. Real-time Hub Configuration tab.
PARAMETER
EXPLANATION
Reference dosimeter, Serial
number
Enter the serial number of the reference dosimeter for use with the system.
Radio range
Nearby
,
In lab
or
Maximum range
. The default setting
In lab
is suitable in
most environments. Use
Maximum range
with care, since it can pick up
dosimeters from other labs.
Unit
Displayed dose and dose rate unit. 1 mSv = 0.1 rem.
Language
The language displayed on the Real-time View, for example in error
messages. The configuration interface is not affected by this setting.
Decimal separator
Use point or comma as decimal separator in the system. For the data export,
Point
as decimal separator will result in comma as list separator, and
Comma
as decimal separator will result in semicolon as list separator.
Browser access
The Real-time View can be accessed from the button Real-time View on the
sign in page, or by pointing a web browser to
https://hubIP/display
Note: More than 3 simultaneous connections to the Real-time View may
impact the refresh rate.
It is possible to restrict access to up to two IP addresses, where the Video
Unit normally should be one of them. Input fields for the IP addresses are
shown when selecting
Only specified IP addresses
.
Clear history and reboot
Clear all the dose history data collected in the hub. The configurations are
not affected, but the information under the
Log
tab is erased.
Service mode
For troubleshooting purposes, you can activate
Service mode
for 20
minutes. Radio signal strength will be displayed for all connected dosimeters,
including the reference dosimeter and other dosimeters configured to be
hidden. Service mode will turn off automatically after 20 minutes. If you want
to turn it off manually, you can do this by rebooting the hub.