3 PRODUCT DESCRIPTION
© 2022 RAYLASE GmbH SUPERSCAN IV
MN105-en / v1.0.12
31
3.7.11
Requirements for cooling air
To avoid contamination of the optical elements and the resulting destruction by the laser beam, the
cooling air must meet the requirements of ISO 8573-1:2010 [1:0(0.05):0(0.005)]:
Pore filter
Pore filter ≤ 5 μm
Required air pressure on the deflection unit
1-1.5 bar
Air flow
approx. 20 l/min
Max. water shares
≤ 0.05 g/m
3
Max. oil shares
≤ 0.005 mg/m
3
Tube outer diameter
4 mm
3.7.12
Requirements for cooling water
To avoid destruction of the aluminium housing by pitting, the cooling water must meet the
requirements listed in the following table.
NOTE
Pitting in aluminium cooling channels
C1 -05 0
Avoid copper in the cooling circuit. Copper causes pitting in the aluminium cooling channels
unless suitable additives are used. Damage caused by pitting is excluded from the warranty.
The mechanical type „S“ is made of stainless steel. Hence there is no pitting even with copper
parts in the cooling circuit.
Cooling water alternatives
Clean drinking water
Deionized water
mixed with 50 % clean drinking water
Deionized water with additives
Pre-mixed cooling fluids (no additives necessary) e.g.:
CCL105 (Ecolab)
Coolflux 42 (Kruckenberg Drucklufttechnik)
Recommended additives
1
Industrial application: TRAC105A_B (Ecolab)
Food industry: Dowcal N (Dow Chemicals)
Recommended cooling temperature
22°C to 28°C
Avoid condensation
Temperature stability
1 °C
Water pressure at the deflection unit
< 3 bar
Water flow and pressure drop
min. 2 l/min, achievable by 0.4 bar pressure drop per
connected unit
Water hardness
< 10 ppm
Recommended pH value
7 to 8.6
Bacterial content
< 1,000 cfu/ml
Tube outer diameter
8 mm
1) Follow the dosage and application instructions of the manufacturer.
3.8
Lifespan
The lifespan of the product is 10 years.
Only for deflection
units with option
"Air Flush"
Only for water
cooled deflection
units