10.2. Learning The System Configuration
The MT5 system keeps a record of the system configuration. The system configuration
defines how many loads and how many channels each load. This configuration
determines which loads and which channels are used. It also controls the channels
available for user settings and operations such as taking of IV curves.
Having the system “learn” the configuration also allows the system to detect when a
load and its data are not available. Data files will contain placeholders for the missing
data and the system will periodically attempt to reconnect to a missing load. For
example, if a load were turned of temporarily, the MT5 controller will automatically
restore a connection once the load is reconnected and turned on.
While it is technically possible for the system to automatically “learn” the configuration,
for example, at power up, Raydec has chosen to require the user to
initiate the “learn”
mode to ensure recording of the proper system configuration.
Raydec delivers new MT5 systems
already configured and it is not necessary to “learn”
a new system. But it is common to add a load to an existing system. In that case, the
system must “learn” the new configuration before this new load operate.
The “learn” mode requires that all of the loads be connected, turned on, and operation
(load status LED blinking)
. This is a very important point. The “learn” mode scans each
load port for a functional load and queries the load for basic information. When learning
the system configuration after attaching a new load, the system will be re-learning the
entire system, not just the new load. So it is critically important that the existing load(s)
be connected and operating normally before re-learning the configuration.
Warning: Be sure ALL loads are connected and turned on and running
before initiating the learn mode.