MT5 Multi-Tracer Operating Manual
page 63
The “Report” button performs the same function as the “Adjust” button, except it doesn’t
change the conversion constant. Its purpose is to make an entry in the calibration
report. This may be useful if it is necessary to document the Multi-Tracer response over
a range of measurement signals.
The “Next” button advances the “Module Voltage Channel” to the next module in the
system. This is a more convenient method to advance to the next module as compared
to using the drop-down list to change module channels. Either before or after changing
to the next module channel, unplug the module connector and move it to the next
channel. Note: Clipping the prongs off of this calibration plug makes it easier to unplug
this connector.
The “Save” button saves all of the changes made to any module-channel, voltage
calibration performed since this dialog was opened. It applies to all changed calibration,
not just the one currently selected. Therefore, it is not necessary to save each module
calibration separately. The normal procedure is to start at module 1, adjust it, use the
“Next” button to advance to the next module, adjust that channel and so on. Once all
module calibrations are completed, all changes can be saved with a single use of the
“Save” button.
Because it is very easy to calibrate module voltages, it is reasonable to save one time at
the completion of module calibration. However, if there is any confusion at any time, and
you want to ensure saving the calibration to that point, use the “Save” button at any time
to prevent inadvertent loss of the new calibration adjustments.
The “Cancel” button cancels all changes made since the dialog was opened or since the
last save and then closes the dialog.
The “Done” button closes the dialog. If adjustments were made and not yet saved, you
will be prompted to do so.