Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 55 of 98
4.4.5.1 Making ICPT settings
In order to be able to make ICPT alarm and threshold settings, it is necessary that a telemetry catheter has
been recognized by the MPR1 and the ICPT is displayed.
further settings
alarm limits
4.4.5.2 Changing ICPT alarm limit settings
The changing of the ICPT threshold settings takes place analogously to the modification of the ICP
threshold settings (see
Changing limiting alarm value settings for invasive pressure measurement).
Parameter
Range with alarm limits
Alarm activation
Press Menu key, if you want
to modify the marked field
4.4.5.2.1 Changing ICPT thresholds
Setting of the ICPT alarm limits as well as the alarm activation/alarm deactivation takes place analogously
to the ICP settings (s.
Changing the pressure limiting value). The option for conventional zeroing
is however not available with ICPT.
4.4.5.2.2 Activating and deactivating ICPT alarms
Alarm setting takes place analogously to setting of the pressure alarm limits (see.
alarm value settings for invasive pressure measurement). Only the monitoring of the pulse amplitude of the
intracranial pressure is not required.
4.4.5.3 ICPT chart
ICPT chart display takes place analogously to pressure chart display (see
4.4.5.4 ICPT trend
ICPT trend display takes place analogously to pressure trend display (see 4.4.3.5