Recommended Liquids 

 
Care should be taken when selecting the most appropriate heating medium for use with the Immersion Heater Circulator to ensure 
optimum performance and user safety. Please refer to the table below as a guide : 
 

Primary Temp Range °C

 

Recommended Medium 

 

Comments 

 

-30 ~ 50  

50% water, 50% laboratory ethylene 
glycol 

Use a lid to reduce the dilution of the medium caused by 
condensing water vapour from the air. 

0 ~ 30 
0 ~ 100 

80% water, 20% laboratory ethylene 
glycol 

Laboratory ethylene glycol will prevent freezing at 0°C and 
reduce evaporation at 100°C and as such is a good all purpose 
heating medium. Use a lid to reduce the dilution of the medium 
caused by condensing water vapour from the air. 

Recommended over water alone for most applications 
between 0° and 100°C

 

5 ~ 60  

Water  

Water alone is suitable for most low temperature applications, 
however laboratory ethylene glycol should be used if working 
above 60°C to reduce evaporation, reduce the chance of 
scalding by steam and help maintain temperature stability. At 
temperatures approaching 100°C the temperature stability 
will be affected due to localised boiling if laboratory ethylene 
glycol is not used . 

The units should not be used to boil 

water. 

 

 
Safety Warnings 

 
Throughout this User Manual, specific warnings will be supplied which relate to the current operation being referred to. These warnings 
are supplied in addition to the main warning labels affixed to the product and the key points outlined in the section of this User Manual 
titled ‘Operator Responsibility – Safety Considerations’. 
 
A graphical symbol as pictured below will be used next to each warning with accompanying text, the danger level for each is described 
below : 
 

 

CAUTION 

Indicates a possibly dangerous situation which may result in serious injury or threat to life as a result of scalding or 
burns if the situation is not avoided. 

 

 

 

CAUTION

 

Indicates a possibly highly dangerous situation which may result in serious injury or threat to life as a result of 
electric shock if the situation is not avoided. 

 

 

 

CAUTION 

Indicates a possibly harmful situation which may result in injury or damage to product or property if the situation is 
not avoided. 

 
Safety Recommendations 

 

The following safety recommendations must be followed to prevent damage or injury. In addition to these safety recommendations, 
it is assumed that the user and operator have had experience in a commercial environment, and had appropriate training in how to 
perform their work safely in accordance with any local operational health and safety regulations. The operator and all users should 
be well versed in local emergency procedures as per the workplace safety regulations in effect. 
 

Содержание TH8500

Страница 1: ...www ratek com au www ratek com au www ratek com au www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Au...

Страница 2: ...the condition of the product to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product...

Страница 3: ...ion 10 Unpacking and Installing 10 Filling The Water Bath 11 Adjusting the Mounting Clamp 12 Adjusting the Pump Flow Rate 12 Attaching a Hose For External Pump Out 13 Connecting Power 13 Switching On...

Страница 4: ...r use in a clean laboratory environment only where adequate ventilation a good power supply and provisions for routine cleaning are available The Immersion Heater Circulator should not be used outdoor...

Страница 5: ...he equipment before using the product Plug the equipment directly into a wall power outlet Do not plug the equipment into a multi socket adapter of any kind The equipment is intended for operation in...

Страница 6: ...es Colours Black on a yellow background The Immersion Heater Circulator is fitted with a removable guard which protects against contact with the heating element This guard MUST REMAIN FITTED at all ti...

Страница 7: ...ollowing operating procedures before installing or operating the Immersion Heater Circulator in order to achieve the best performance and maximum attainable user safety Identification of Controls Func...

Страница 8: ...il water Safety Warnings Throughout this User Manual specific warnings will be supplied which relate to the current operation being referred to These warnings are supplied in addition to the main warn...

Страница 9: ...serious injury or danger to life CAUTION Always work above the level of the Immersion Heater Circulator to avoid spillage of dangerously hot water At no point should the user operator animal or any ot...

Страница 10: ...ath to allow for adequate ventilation and for steam to dissipate Ensure that there is a suitable mains power supply outlet within reach of the supplied mains power lead without placing any strain what...

Страница 11: ...ur Immersion Heater Circulator before filling the water bath if you are unsure of your water supply CAUTION Before filling the water bath ensure the mains power supply outlet is switched off and that...

Страница 12: ...th as required Once the depth has been set re tighten the 2 clamp screws by gently tightening in a clockwise direction Step 2 Once the unit is placed on the bath gently tighten the thumb screw by turn...

Страница 13: ...diameter no larger than 15mm Silicon hose is recommended for most applications Simply push the hose over either outlet until it touches the guard a hose clamp may be used if required however this is...

Страница 14: ...s power supply outlet Switching On The Immersion Heater Circulator CAUTION Ensure you have filled the water bath to at least the minimum required level before switching on power Failure to do so may c...

Страница 15: ...am and splashing of water when the Immersion Heater Circulator is operating at these high temperatures particularly whilst loading and unloading containers racks or accessories ALWAYS use appropriate...

Страница 16: ...ow You should choose an appropriate profile number based on the maximum number of segments your profile will require Profile Numbers Maximum Number of Segments 1 thru 4 16 5 thru 7 32 8 and 9 64 To co...

Страница 17: ...ts enter the number of times the profile should be repeated cycled The target set point in C for a ramp type segment Time duration or ramp rate for a ramp type segment entered in units specified at Al...

Страница 18: ...elected segment Note Press simultaneously to move backwards through the profile parameters Once you finished entering your profile parameters push to exit the profile menu and return to the normal sta...

Страница 19: ...le by the user The alarms are configured as deviation alarms whereby they operate relative to the desired set point If the current temperature deviates from the set temperature by the margin configure...

Страница 20: ...uch as silicon insulated tongs to slowly remove the items and place them onto a water proof and heat proof surface Continue to load or unload your containers racks or accessories until complete Switch...

Страница 21: ...eets the environmental conditions required as detailed in the technical specifications of this User Manual Turn off the Mains Power Switch by setting it to the position marked with a circle Unplug the...

Страница 22: ...make any part of the control panel any exposed control or receptacle or any part of the Immersion Heater Circulator excessively wet If these receptacles and controls remain wet once electrical power...

Страница 23: ...s Power Switch Unplug the equipment from the mains power supply outlet and remove it from the water bath Empty the water bath Fill the water bath with enough cold water to just cover the top coil of t...

Страница 24: ...used an unpleasant smell may also be noticed as a result of the impurities in the water becoming hot This is normal Once the test has been completed turn off power to the Immersion Heater Circulator...

Страница 25: ...use Altitude up to 2 000 Metres Temperature 5 Celsius to 40 Celsius Ambient temperature will limit the minimum achievable water temperature Maximum relative humidity 80 for temperatures up to 31 Cels...

Страница 26: ...the contact details provided in the section of this User Manual titled Ratek Service Contact Information if you believe your equipment is exhibiting a fault condition Warranty Conditions This Ratek pr...

Страница 27: ...der the Warranty you must present the product together with proof of purchase or issue to the store where you purchased the product from If the product is defective and does meet the Warranty you will...

Страница 28: ...ammable Immersion Heater Circulator User Manual Page 28 of 28 Provide a full and complete fault description and your return contact details in the package and return the product as advised by the serv...

Отзывы: