you to select peaks (typically within the white window) that you know are correct but are just below the auto
calibration peak height requirements.
The manual calibration routine enables the user to calibrate on any substance in the library, as long as the vapor
material of that substance is available to collect a sample.
Dual Peak Calibrate - Vapor Mode
1. Remove the desorber, and install the vapor plate.
see "Removing the Desorber and Installing the Vapor Plate" on page 94)
2. Select Menu > Manual Calibrate.
Figure 3-14
Vapor: Calibrate Options Dialog
3. Press Dual Peak Calibrate.
A message to select the calibration substance displays.
Figure 3-15
Vapor: Select a substance to calibrate on (neg)
4. Select a substance from the negative ion list of substances.
•
If you want to calibrate using the internal calibrant, select
Vapor-Neg-CAL
.
•
If you would like to calibrate using the Check Sample vial, select
Vapor-Neg-Ver
.
36
Rapiscan Systems Proprietary and Confidential
CHAPTER 3 │ User's Guide