Operating Instructions RS cure
│
D cure
│
P cure
│
SHERAprint cure
│
3Decure
Rapid Shape GmbH
│
DOCR000251
24
│
27
microfibre cloth moistened with glass cleaner or isopropanol.
For persistent soiling, e.g. hard incrustations, the glass bottom
must be replaced. Contact Customer Service for more
information.
NOTE!
Wearing protective gloves protects the skin from contact
with skin damaging materials. It is therefore essential to
observe the information provided by the material
manufacturer.
NOTE!
Never spray the cleaning medium directly into the
exposure chamber. This can damage the LEDs and
other components. When cleaning, always use
moistened microfibre cloths with the cleaning medium.
For proper operation of the post curing unit, the
exposure chamber must always be kept clean and dust-
free.
8.2.2
Check & Clean Filter
Required accessories:
Personnel:
Vacuum cleaner
Operator
For checking and vacuuming the filter cartridges in the machine,
please proceed as follows:
1.
Make sure the exposure chamber is empty.
2.
Open the lid. The filter cartridges of the upper part of the
machine are located in the front panel (below the touch display).
You can see it through the slot openings.
3.
Move the vacuum cleaner
’s
suction tube along the slot
openings. Then close the lid again.
4.
Pull the machine either 3-5 cm over the edge of the table or tilt
the machine slightly backwards. The filter cartridges of the
machine
’s
lower part are located in the front panel.
5.
Move the vacuum cleaner
’s
suction tube along the slot
openings.
6.
Then gently put the machine back.
8.2.3
Calibrate the LEDs
Required accessories:
Personnel:
ACCS Sensor
Operator
For calibrating the LED, please proceed as follows:
1.
Switch on the machine.
→
Chapter 6.2
2.
Press the
‚Settings’
button. The settings are opened.
3.
Press the
‚Calibration’
button. The calibration menu is opened.
4.
Follow the instructions on the display.