Preparing the Extension Unit for Use
R&S
®
ZVAX-TRM
14
User Manual 1325.1268.02 ─ 04
ESD Protective Measures
To protect the extension unit against damage due to Electrostatic Discharge (ESD) use
the wrist strap and grounding cord supplied with the network analyzer and connect
yourself to the GND connector at the front panel of the R&S
ZVA or R&S
ZVT. For
details refer to the Quick Start Guide of your analyzer.
USB Connection to the Analyzer
●
The extension unit is intended for direct connection to R&S
ZVA or R&S
ZVT net-
work analyzers according to the instructions in
Chapter 3.4.5, "Connecting the
on page 22 and in the analyzer’s help sys-
tem. Do not connect the unit to other USB hosts, e.g. a PC, and do not insert any
USB hubs between the analyzer and the unit, as this may cause damage to the
unit or the host.
●
It is safe to connect or disconnect the extension unit while the network analyzer is
operating. Never connect the unit during a firmware update.
3.2 Front Panel Tour
The front panel of the extension unit provides the standby key, two USB connectors
(type
A), RF connectors for ports 1, 2, 3, and 4, as well as several indicator LEDs.
Figure 3-1: R&S
ZVAX-TRM Front Panel
3.2.1 Standby Key
The standby key connects/disconnects all internal modules of the extension unit to/
from the DC supply voltages generated by its internal power supply.
The two LEDs indicate whether the extension unit is ready to operate (green LED on)
or in standby state (yellow LED on).
Front Panel Tour