Preparing the Analyzer for Use
R&S
®
ZVA
15
Getting Started 1145.1090.62 ─ 13
Input signals
The maximum RF input levels at all SMA inputs according to the front panel labeling or
the data sheet must not be exceeded.
In addition, it is important that the signal fed in at the SMA inputs contains no DC off-
set, as this may impair the measurements and even cause damage to the instrument.
1.1.8 Additional Hardware Options
The following hardware options can be mounted on the right of the NAVIGATION and
SUPPORT keypads:
●
Option R&S
ZVA-B18, Removable Hard Disk, replaces the internal hard disk by a
removable compact flash card. The compact flash card can be inserted at the front
panel of the instrument. To ensure failure-free operation, avoid placing external
cables close to the compact flash card.
●
Option R&S
ZVA-B8, Converter Control, provides output connectors to control the
output power of a frequency converter with external attenuators, R&S
ZVA-ZxxxE.
This option is available for R&S
ZVA network analyzers with an upper frequency
limit of 20 GHz or higher (R&S
ZVA 24, R&S
ZVA40 ...). For information on mea-
surements with external attenuators, refer to the network analyzer's help system
and to the Getting Started guide R&S
ZVA-ZxxxE, stock no. 1307.7197.62.
Contact your R&S service representative, if you wish to obtain and install one of the
additional hardware options.
1.2 Rear Panel Tour
This section gives an overview of the rear panel controls and connectors of the net-
work analyzer.
Rear Panel Tour