Data management
R&S
®
ZNL/ZNLE
95
User Manual 1178.5966.02 ─ 19
After you use the [PRESET] function, the history of previous actions is deleted, i.e. any
actions performed previously cannot be undone or redone using the [UNDO/REDO]
keys.
Remote command:
*RST
or
SYSTem:PRESet
To restore the default configuration for a single channel setup
The default measurement settings can also be reset for an individual channel setup
only, rather than resetting the entire instrument.
► In the "Overview", select the "Preset Channel" button.
The factory default settings are restored to the current channel setup. Note that a
user-defined recall settings file is
NOT
restored.
●
The procedure above applies to spectrum mode only (n.a. for R&S
ZNLE).
●
For the R&S
ZNL/ZNLE a single channel setup can only be reset via remote con-
trol.
Remote command:
6.2
Storing and recalling instrument settings and mea-
surement data
Access
: "Save"/ "Open" icon in the toolbar
Or
: [FILE]
Possibly you would like to restore or repeat a measurement you performed under spe-
cific conditions on the instrument. Or you want to evaluate imported data in another
application on the R&S
ZNL/ZNLE and would like to restore the measurement settings
applied during measurement. In these cases, you can store and recall instrument and
measurement settings, and possibly other related measurement data.
Two different methods are available for managing instrument settings:
●
Quick Save/Quick Recall - a defined set of instrument settings or channel setups
are stored or recalled quickly in just one step
●
Configurable Save/Recall - a user-defined set of instrument settings or channel
setups are stored to a definable storage location
Storing and recalling instrument settings and measurement data