Command reference
R&S
®
ZNL/ZNLE
907
User Manual 1178.5966.02 ─ 19
<CalUnitPort1>
Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2>
Second test port number
*RST:
n/a
Example:
CORR:COLL:AUTO:PORT FNPort, '', 1, 2, 2, 4, 4,
1
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and ports
2, 4, and 1 of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:REPeat
Use this command to reuse measurement data of the active automatic calibration (if
available) for the next automatic calibration. This data reuse allows an automatic cali-
bration to be repeated without repeating all measurements.
Reusing existing calibration data is particularly useful for
OSM measurements can be repeated without reconnections, but whose unknown
through measurements can not.
Note
that this command does not work with "single-command calibrations" performed
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
.
Suffix:
<Ch>
.
Channel number
Example:
See
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
<Boolean>
Available if the calibration unit selected using
(ICS) only. Applies to all inline calibra-
tion units (ICUs) that are connected to the selected ICS and that are used with factory
characterization data .
If set to ON (1), the R&S
ZNL/ZNLE firmware adjusts the factory characterization data
according to the ICU temperatures.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Manual operation:
See
VNA command reference