<Mk>
This numeric suffix is ignored and may be set to any value
because the bandfilter search functions always use markers M1
to M4.
Parameters:
<Bandwidth>
Difference between the band edges and the center response
value of a bandfilter peak; must be negative for a bandpass
search and positive for a bandstop search.
Range:
For bandpass: -100.00 dB to -0.01 dB; for band-
stop: +0.01 dB to +100.00 dB
Increment: 0.03 dB
*RST:
-3 dB
Default unit: dB
Example:
CALC:MARK:FUNC:BWID:MODE BST
Select a bandstop filter search.
CALC:MARK:FUNC:EXEC BFIL
Initiate the bandpass filter search for the current trace. Create
markers M1 to M4.
CALC:MARK:SEAR:BFIL:RES ON
Display the marker info field in the diaram area.
CALC:MARK:BWID 6
Select a 6-dB bandwidth for the bandstop.
CALC:MARK:BWID?
Query the results of the bandfilter search. An error message is
generated if the bandfilter search fails so that no valid results are
available.
CALC:MARK:SEAR:BFIL:RES:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
Manual operation:
See
CALCulate:MARKer:COUPled[:STATe]
<Boolean>
Enables marker coupling to the active trace of the active channel or disables it.
Use
to select the suitable coupling type before
setting
CALCulate:MARKer:COUPled:STATe
to
ON
.
Parameters:
<Boolean>
ON | OFF - enables or disables marker coupling.
*RST:
OFF
VNA command reference