Command reference
R&S
®
ZNL/ZNLE
678
User Manual 1178.5966.02 ─ 19
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:INTerpol
<LimLineInterpolation>
"Interpolation" determines whether the limit line segment is interpolated linearly or log-
arithmically. The latter may be more suitable for logarithmic sweeps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<LimLineInterpolation>LOGarithmic | LINear
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt
<FreqPowTime>
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP
<StimVal>
These commands change the start and stop stimulus values (i.e. the smallest and the
largest stimulus values) of a limit line segment. A segment must be created first to
enable the commands (e.g
ter 9.4.1.1, "Rules for limit line definition"
Tip:
To define the stimulus values of several limit line segments with a single com-
CALCulate<Chn>:LIMit:CONTrol[:DATA]
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<StimVal>
Stimulus value confining the limit line segment.
If specified, the unit of a stimulus value must in accordance with
the sweep type of the active channel (
Hz
for frequency sweeps,
dBm
for
power sweeps (R&S
ZNL only), and
s
for time sweeps. For CW
mode sweeps, stimulus values are dimensionless.
Default unit: NN