Getting Started
R&S
®
ZNL/ZNLE
47
User Manual 1178.5966.02 ─ 19
4.2.2.2
Li-Ion battery packs and DC power connector
The Li-Ion battery pack and DC power connector are available as an optional acces-
sory (R&S
FPL1-B31 and R&S
FPL1-B30).
With the Li-ion battery pack, the R&S
ZNL can be operated independently of an AC or
DC power supply. The instrument can house 2 Li-ion batteries which can be charged
both via AC or DC power supply.
For safety information concerning batteries, see
As an alternative, a DC power supply connector is available. DC power supplies from
+12
V to +24
V and from 13
A to 6.5
A can be used. Connect the connector according
to the following diagram:
Pin
Description
1
2
3
1
Plus
2
Ground
3
Not used
These options are not available for the R&S
ZNLE.
If you use an external power supply unit to supply safety extra-low DC voltage (SELV)
to the instrument, be sure to meet the requirements for reinforced/double insulation in
accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1) or DIN/EN/IEC
60950 (UL 1950, CSA C22.2 No. 950). Provide current limitation in accordance with
DIN EN 61010-1 Appendix F2.1.
Also see
4.2.2.3
GPIB interface
The GPIB interface is available as an optional feature (R&S
FPL1-B10).
The optional GPIB interface is in compliance with IEEE488 and SCPI. A computer for
remote control can be connected via this interface. To set up the connection, a shiel-
ded cable is recommended.
For details, see "Interfaces and Connectors" in the appendix of the user manual or
online help.
For details, see
Chapter 14.2.2, "GPIB interface"
Instrument tour