Network and remote settings
R&S
®
ZNL/ZNLE
174
User Manual 1178.5966.02 ─ 19
6. Select "Done" to close the dialog box.
The drive is shared and can be accessed by the selected users.
8.5.5
How to control the R&S
ZNL/ZNLE via the web browser interface
Via the LAN web browser interface to the R&S
ZNL/ZNLE, one or more users can con-
trol the instrument remotely from another PC without additional installation. Most instru-
ment controls are available via the front panel simulation. File upload and download
between the instrument and the remote PC is also available.
To access the R&S
ZNL/ZNLE via the web browser interface
1. Start a web browser that supports html5 (W3C compliant).
2. Enter the IP address of the R&S
ZNL/ZNLE in the browser's address bar.
The R&S
ZNL/ZNLE's Welcome page is displayed.
3. In the navigation pane, select "Instrument Control > Web Control".
The instrument's display is shown in a new browser window, with a software front
panel displayed beside or below it.
4. Use the mouse cursor to access the functionality in the software front panel or in
the display as you would directly on the instrument's front panel.
To exchange files with the R&S
ZNL/ZNLE
You can download files, for example stored measurement data, from the R&S
ZNL/
ZNLE to the remote PC, or upload files, for example limit line definitions, from the PC
to the R&S
ZNL/ZNLE.
1. In the web browser, select the Welcome page window.
2. In the navigation pane, select "Instrument Control" > "File Upload" or "File Down-
load".
How to set up a network and remote control