Remote Commands
R&S
®
ZNH
243
User Manual 1334.5985.02 ─ 02
Parameters:
<RefPosition>
Numeric value in the range from 0 to 10.
*RST:
10
Example:
DISP:LOSS:REF:POS 5
Moves the reference to the fifth grid line from the bottom.
DISPlay<1...2>:LOSS:Y:SCALe
<DisplayRange>
This command defines the display range of the vertical axis for the cable loss mea-
surement format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1 dB to 100 dB.
*RST:
100 dB
Default unit: dB
Example:
DISP:LOSS:Y:SCAL 20
Defines a display range of 20 dB
DISPlay<1...2>:MAGNitude:REFerence
<RefLevel>
This command defines the reference value for the magnitude measurement format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<RefLevel>
Numeric value in the range from -80 dB to 30 dB
*RST:
0 dB
Default unit: dB
Example:
DISP:MAGN:REF -10
Defines a reference level of -10 dB.
DISPlay<1...2>:MAGNitude:REFerence:POSition
<RefPosition>
This command defines the position of the reference value in the diagram for the magni-
tude measurement format.
Each step shifts the reference position one grid line up or down.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Remote Control - Commands