Remote Commands
R&S
®
ZNH
229
User Manual 1334.5985.02 ─ 02
[SENSe:]CORRection:CDATa?
<ErrorTerm>, <SourcePort>, <LoadPort>
This command reads system error correction data for calibration method and port com-
bination <SourcePort>, <LoadPort>.
Calibration type
Calibration method
Available error terms
One-port normalization (reflection)
using an open or a short standard
"Refl Norm Open" and "Refl Norm
Short"
"REFLtrack"
Full one port
"Refl OSM"
"TRANSTRACK"
One-port, two-port normalization
"Trans Norm Both"
"DIRECTIVITY", "SRCMATCH",
"REFLTRACK", "LOADMATCH",
"TRANSTRACK"
Parameters:
<ErrorTerm>
DIRectivity | SRCMatch | REFLtrack | LOADmatch | TRANstrack
String parameters describing the different error terms, depend-
ing on the current calibration method. Each term contains one
complex value (real and imaginary part) for each sweep point.
The parameter must be transferred in full length.
DIRectivity
Directivity at the <SourcePort>
SRCMatch
Source match at the <SourcePort>
REFLtrack
Reflection tracking at the <SourcePort>
LOADmatch
Load match at the <LoadPort>
TRANstrack
Transmission tracking between the <SourcePort> and the
<LoadPort>
Range:
The error terms are dimensionless complex num-
bers.
<SourcePort>
Source port number.
<LoadPort>
Load port number. If the error term is not related to the load port,
a dummy number can be used; e.g.
CORR:CDAT
'REFLtrack', 1, 0
Usage:
Query only
Selecting the Result Display
The following commands select the result display.
List of commands
●
MEASurement<1...2>:FUNCtion:SELect