Annexes
R&S
®
ZNC
749
User Manual 1173.9557.02 ─ 13
The default configuration of the signals is listed below. Notice that
*RST
or "Preset" do
not change the configuration of the Universal Interface. Use
to restore default values.
Signal
*RST Configuration
Port A, Port B
Output ports, all bits "High" (decimal 0)
Port C, Port D
Input ports
/PORT C STATUS, /PORT D STATUS
"Low"
/OUTPUT 1, /OUTPUT 2
"High"
/SWEEP END
"High"
/PASS_FAIL
"High"
SCPI commands:
(resets all configurable signals)
9.1.4.4
Timing of Control Signals
The timing of the essential measurement control signals is shown in the figure below.
The duration of the shaded time intervals depends on the measurement settings.
Fig. 9-8: Timing of measurement control signals
The figure above corresponds to the default configuration of the /PASS FAIL signal. For
alternative configurations refer to the documentation of the
CONTrol:HANDler:PASSfail...
commands in
, on page 460. The figure contains the following pulse durations and response
times.