Command Reference
R&S
®
ZNC
637
User Manual 1173.9557.02 ─ 13
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DEFine<ListNo>
<Frequency>, <TransCoeff>
Adds a new frequency and transmission coefficient to the end of the power loss list. The
query returns the frequency and transmission coefficient no.
<ListNo>
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo>
Number of point in the list. This suffix is only used for queries.
Parameters:
<Frequency>
Stimulus frequency value. If several points with identical frequen-
cies are added, the analyzer automatically ensures a frequency
spacing of 1 Hz.
Range:
Stimulus values outside the frequency range of the
analyzer are allowed.
*RST:
n/a
Default unit: Hz
<TransCoeff>
Transmission coefficient
Range:
-300 dB to +200 dB
*RST:
n/a
Default unit: dB
Example:
See
Manual operation:
See
"Transmission Coefficients; Insert / Delete / Delete All"
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>:ALL
Clears the power loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo>
Number of point in the list. This suffix is ignored; the command
deletes all points.
Example:
See
Usage:
Event
SCPI Command Reference