Command Reference
R&S
®
ZNC
615
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]SWEep:POINts
<SweepPoint>
Defines the total number of measurement points per sweep ("Number of Points").
Values between 1 and 5,001 can be set.
Suffix:
<Ch>
.
Channel number 1
Parameters:
<SweepPoint>
Number of points per sweep
Range:
see above
*RST:
201
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation:
See
"Start / Stop / Center / Span Frequency (Power / Time)"
[SENSe<Chn>:]SWEep:SRCPort
<Port>
Selects a source port for the stimulus signal. The setting acts on the active trace. The
effect of the drive port selection depends on the measurement parameter associated to
the active trace:
●
If an S-parameter S
<out><in>
is measured, the second port number index
<in>
(input
port of the DUT = drive port of the analyzer) is set equal to the selected drive port:
Drive port selection affects the measured quantity.
●
If a wave quantity or a ratio is measured, the drive port is independent of the measured
quantity:
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Port>
Test port number of the analyzer, 1 to 2, 3 or 4, depending on the
analyzer model.
*RST:
1 (test port 1)
SCPI Command Reference